Measurement Specialties
KPSI
®
Digital Series User’s Manual
Page 25
www.Meas-Spec.com
XT
Set Averaging
Time
aXTt!
“t”= +
averaging time
in
seconds (0 to 25)
:
:
aD0!
a0011
:
a (
Service Request
response)
Service Request
resp. returned in
001 seconds.
“1” is
number values
in “
D0
” resp.:
as
“s” = number samples to be
averaged during selected time
XUM
Set /Read
Measurement
Mode of M/M0,
C/C0, MC/MC0,
CC/CC0
Commands
Drop optional
argument [m] to
Read mode
aXUM[m]!
m=missing to read mode
or
m=+0 to set mode for
pressure data
only
(default) or
m=+1 to set mode for
pressure data
and
temperature data
just like
M7/C7/MC7/CC7
commands
:
aD0!
attt1
:
a {
Service Request
response)
Service request
resp. returned in m
seconds. 1 is
number
values
in
D0
resp.
am m=+0 (Press.only) or
m=+1(Press. and Temp)
XUP
Set Pressure
Units
aXUPn[d]!
n=+0 feet of water
n=+1 psi
n=+2 kPa
n=+3 cm of water
n=+4 m of water
n=+5 mm of water
n=+9 user units
[+d=optional number of
places to right of the
decimal point]
:
aD0!
a0012
:
a (
Service Request
response)
Service Request
resp. returned in
001 seconds.
“2” is
number values
in“
D0
” resp:
NOTE:
Make sure you use
XUU
command to set the slope and offset
for the desired user units.
aud
“u” is
+units
selected.
“d” is +
number of digits to right of
decimal point
.