3.6 Measurement Range 

A measurement range determines the measuring limits. The meter has several measurement ranges. 

1)

 

Select a proper measurement range: 

 

If the selected measurement range is too small (the testing signal exceeds the measurement range), the 

meter will display “OL” on the LCD. 

 

If the selected measurement range is too large, the testing results may have a relatively low resolution. 

2)

 

Auto range and Manual range 

The meter has both auto range and manual range for the measurements. 

 

In the AUTO measuring mode, the meter will automatically select a most suitable range for the testing 

signal. In AUTO mode, the meter will display “AUTO” on the LCD. 

 

In  the  MANUAL  measuring  mode,  the  measuring  range  is  selected  manually  by  pressing  RANGE 

button. 

 

If  there  is  more  than  one  range  in  a  measuring  function,  the  meter  will  default  in  AUTO  range.  This 

allows users to override the auto range and lock the meter in a specific range. 

 

In HOLD or MAX/MIN recording mode, changing the measurement range manually will automatically 

exit the HOLD or MAX/MIN mode. 

3)

 

Enter or exit mauran range mode 

 

Press RANGE button to select manual range mode. Then, every press on the RANGE button, the meter 

steps to a higher range and will return to the lowest range when the higher range is stepped into.   

 

Press RANGE button for at least two seconds can also return to auto range mode.   

 

 

4. OPERATION INSTRUCTIONS 

4.1 DC Voltage Measurement (DCV) 

The  measurement  of  DC  voltage  has  three  modes:  DCV, 

DCV+ACV and dBm. 

 

Set the function key to “

” position. 

 

Press SELECT button to select measurement mode. 

 

According  to  practical  demands,  press  REL

,  MAX/MIN 

and  2nd  VIEW  buttons  to  have  relative  measuring  or 

record. 

 

Connect the black test lead to “COM” terminal and the read 

test lead to “V Ω Hz” terminal. 

 

The meter is in auto range mode and will automatically 

select a most suitable range. To select a range manually for 

DCV measurement, press RANGE button to select a proper range. There are four ranges to choose from: 

8.0000V/80.000V/800.00V/1000.0V 

 

Touch the test leads to the test points and read the display on both the primary and secondary display. 

 

Press SELECT 

Primary display  Secondary display 

Press 2nd VIEW 

DCV 

DCV 

 

DCV+ACV 

DCV 

 

dBm 

dBm 

Hz/ACV/DCV/DCV+ACV 

 

 

Содержание MT8145

Страница 1: ...Benchtop Instrument 80000 Counts 4 7 8 Bench Digital Multimeter Operation Manual...

Страница 2: ...s read this operation manual carefully before using this meter and follow the following safety guidelines Before use check and make sure that the instrument s plastic chassis the test leads and the in...

Страница 3: ...2 2 PANEL INSTRUCTIONS Fig 2 1 Front panel Fig 2 2 Rear Panel...

Страница 4: ...s TIMER key to start the secondary display for counting time Press TIMER key again to turn off the counting time display When the secondary display is display counting time press key SELECT to enter B...

Страница 5: ...ppears on the LCD display The relative measuring function measures the difference between the testing value and the reference value The current readings on the secondary display are used as a relative...

Страница 6: ...C DC AC DC AC voltage or current 10 PH PH Peak Hold Peak Hold 11 A H Auto Hold 12 AVG Average reading 13 Auto Auto mode 14 APO Auto power off sign 15 RS232 Communication on annunciation 16 MAX MIN MAX...

Страница 7: ...alue on the primary display and on the secondary display c If the measuring value is between and upper and lower limit the LCD displays the measuring value on the primary display and on the secondary...

Страница 8: ...lly will automatically exit the HOLD or MAX MIN mode 3 Enter or exit mauran range mode Press RANGE button to select manual range mode Then every press on the RANGE button the meter steps to a higher r...

Страница 9: ...choose from 8 0000V 80 000V 750 00V Connect the black test lead to COM terminal and the read test lead to V Hz terminal Touch the test leads to the test points and read the display on both the primary...

Страница 10: ...mpt an in circuit current measurement when the fuse is blown or when the voltage between open circuit and the ground is 1000V To avoid damage to the meter check the meter s fuse before proceeding Befo...

Страница 11: ...rations apply the same as explained in section 4 4 Press SELECT Primary display Secondary display Press 2nd VIEW DCA DCA ACA ACA DCA ACA DC AC A ACA ACA Hz ACA Hz 4 6 Resistance Measurement WANRNING T...

Страница 12: ...ance Use the DC function to confirm that the capacitor is discharged NOTE Some capacitors have polarities In measuring polarities capacitors touch the red test lead to the positive polarity and the bl...

Страница 13: ...nd read the display Measurement mode Primary display Secondary display Press 2nd VIEW Hz Hz kHz MHz High frequency mode Hi Hz In Hi Hz mode the frequency testing range is 8MHz 1000MHz divided into fou...

Страница 14: ...Read the display 4 10 Diode and Continuity Check WANRNING To avoid damage to the meter or the testing equipments disconnect circuit power and discharge all high voltage capacitors before measuring Us...

Страница 15: ...software which allows easy connections to PC for PC control Software operating environment Windos9x NT 2000 XP Software version V1 0 5 1 Installation Plug the CD disk into the CD drive Start up the se...

Страница 16: ...es you an easier view 5 11 Push Button Panel Basically the functions of the buttons on this panel are the same as the buttons on the multimeter Double click on the function buttons is in the same way...

Страница 17: ...m part of the panel indicate the current status of the data recorder Battery indicates the power supply status of the multimeter If power supply is in good status it will display Battery High Otherwis...

Страница 18: ...20A 1mA Table 4 ACA True RMS Range Resolution Accuracy Remarks 80mA 1 A 50Hz 5kHz 0 2 rdg 10 Fuse F750Ma 250v F13A 250V Voltage drop 800mV Max input current 20A up to15 seconds 800mA 10 A 8A 0 1mA 50H...

Страница 19: ...nd ground 1000V RMS Continuity beeper Approx 3kHz Display dual display 80000 update 4 times Sec Bar graph 23 segments update 40 times Sec Electromagnetic compatibility For all ranges and functions exc...

Страница 20: ...ter 7 1 Fuse replacement NOTE Please select fuse of the same specification to replace it Power fuse 200mA 250V located in input power socket Fuse for measuring current 800mA 250V located in the curren...