- 6 -

6. AC110V/220V selector switch 

7. Fuse 

8. Power plug 

9. RS232 interface 

 

 

3. SPECIAL FUNCTIONS 

3.1 Auto Power Off 

The meter has two 

powers

 off function. In normal conditions when the meter is power on, if there is no operation 

on any of the buttons or function keys within 30 mints, the meter will be automatically powered off. Five mints 

before auto power off, the audible five beepers give indication that the meter is gong to be power off. 

 

3.2 Setting Upper and Lower Limits for Measurement 

The upper and lower measurement limits can be set up in the following steps: 

1.

 

Set the upper  limit:  power on  the meter 

  select  range 

  press SET  button 

  press SELECT  button  to 

start the upper setting mode (“

” appears on secondary display) 

  press 

  buttons to adjust 

the value 

  press SET button to validate the new input. 

2.

 

Set the  lower  limit:  power on  the  meter 

  select  range 

  press SET  button 

  press SELECT  button  to 

start the lower setting mode (“

” appears on secondary display) 

  press 

  buttons to adjust 

the value 

  press SET button to validate the new input. 

3.

 

After setting up the upper and lower limits, measurement can be carried out and the LCD will have 3 kinds of 

displays to the results: 

a)

 

If the  measuring value exceeds the  upper limit,  the  LCD  displays the  measuring value  on  the primary 

display and “

” on the secondary display. 

b)

 

If  the  measuring  value  exceeds  the  lower  limit,  the  LCD  display  the  measuring  value  on  the  primary 

display and “

” on the secondary display. 

c)

 

If the measuring value is between and upper and lower limit, the LCD displays the measuring value on 

the primary display and “

” on the secondary display. 

 

3.3 Set up the Time for Measurement 

1.

 

Press TIMER button to set enter the setup of time. The secondary display displays “0.00.00” 

2.

 

Press SELECT button and the last digit of “0.00.00” on the secondary display glitters, which means this digit 

can be adjusted. Use the 

  buttons to adjust the digit value. (The first digit from the left is hour, the 

second and third are minute, the fourth and fifth are second.) 

3.

 

Press TIMER button to validate the new setup. By now the measuring time is set. 

 

3.4 Analog Bar Graph 

The function of analog bar graph is imitating the analog needle of the meter but without the overshoot. The bar 

graph refreshes data 40 times per second. Because the graph responds 10 times faster than the digital display, it is 

widely used in the application of peak value test and zero calibration. It is also commonly used in observing the 

rapid  variation  of  input  signals.  The  bar  graph  has  23  segments.  The  number  of  lit  segments  is  relative  to  the 

full-scale  value  of  the  selected  range.  One  unit  of  the  bar  graph  represents  4000  counts/bar  except  when  in  the 

relative mode. The polarity is indicated at the left of the bar graph. 

 

3.5 Square Wave Output 

The  square wave output  is  a  very  useful  function.  With adjustable square wave frequency  and duty  cycle, users 

can carry out the pulse wave modulation (PWM) output, and adjust the voltage control, timer control and clock 

Содержание MT8145

Страница 1: ...Benchtop Instrument 80000 Counts 4 7 8 Bench Digital Multimeter Operation Manual...

Страница 2: ...s read this operation manual carefully before using this meter and follow the following safety guidelines Before use check and make sure that the instrument s plastic chassis the test leads and the in...

Страница 3: ...2 2 PANEL INSTRUCTIONS Fig 2 1 Front panel Fig 2 2 Rear Panel...

Страница 4: ...s TIMER key to start the secondary display for counting time Press TIMER key again to turn off the counting time display When the secondary display is display counting time press key SELECT to enter B...

Страница 5: ...ppears on the LCD display The relative measuring function measures the difference between the testing value and the reference value The current readings on the secondary display are used as a relative...

Страница 6: ...C DC AC DC AC voltage or current 10 PH PH Peak Hold Peak Hold 11 A H Auto Hold 12 AVG Average reading 13 Auto Auto mode 14 APO Auto power off sign 15 RS232 Communication on annunciation 16 MAX MIN MAX...

Страница 7: ...alue on the primary display and on the secondary display c If the measuring value is between and upper and lower limit the LCD displays the measuring value on the primary display and on the secondary...

Страница 8: ...lly will automatically exit the HOLD or MAX MIN mode 3 Enter or exit mauran range mode Press RANGE button to select manual range mode Then every press on the RANGE button the meter steps to a higher r...

Страница 9: ...choose from 8 0000V 80 000V 750 00V Connect the black test lead to COM terminal and the read test lead to V Hz terminal Touch the test leads to the test points and read the display on both the primary...

Страница 10: ...mpt an in circuit current measurement when the fuse is blown or when the voltage between open circuit and the ground is 1000V To avoid damage to the meter check the meter s fuse before proceeding Befo...

Страница 11: ...rations apply the same as explained in section 4 4 Press SELECT Primary display Secondary display Press 2nd VIEW DCA DCA ACA ACA DCA ACA DC AC A ACA ACA Hz ACA Hz 4 6 Resistance Measurement WANRNING T...

Страница 12: ...ance Use the DC function to confirm that the capacitor is discharged NOTE Some capacitors have polarities In measuring polarities capacitors touch the red test lead to the positive polarity and the bl...

Страница 13: ...nd read the display Measurement mode Primary display Secondary display Press 2nd VIEW Hz Hz kHz MHz High frequency mode Hi Hz In Hi Hz mode the frequency testing range is 8MHz 1000MHz divided into fou...

Страница 14: ...Read the display 4 10 Diode and Continuity Check WANRNING To avoid damage to the meter or the testing equipments disconnect circuit power and discharge all high voltage capacitors before measuring Us...

Страница 15: ...software which allows easy connections to PC for PC control Software operating environment Windos9x NT 2000 XP Software version V1 0 5 1 Installation Plug the CD disk into the CD drive Start up the se...

Страница 16: ...es you an easier view 5 11 Push Button Panel Basically the functions of the buttons on this panel are the same as the buttons on the multimeter Double click on the function buttons is in the same way...

Страница 17: ...m part of the panel indicate the current status of the data recorder Battery indicates the power supply status of the multimeter If power supply is in good status it will display Battery High Otherwis...

Страница 18: ...20A 1mA Table 4 ACA True RMS Range Resolution Accuracy Remarks 80mA 1 A 50Hz 5kHz 0 2 rdg 10 Fuse F750Ma 250v F13A 250V Voltage drop 800mV Max input current 20A up to15 seconds 800mA 10 A 8A 0 1mA 50H...

Страница 19: ...nd ground 1000V RMS Continuity beeper Approx 3kHz Display dual display 80000 update 4 times Sec Bar graph 23 segments update 40 times Sec Electromagnetic compatibility For all ranges and functions exc...

Страница 20: ...ter 7 1 Fuse replacement NOTE Please select fuse of the same specification to replace it Power fuse 200mA 250V located in input power socket Fuse for measuring current 800mA 250V located in the curren...

Страница 21: ...oes not work properly take actions as described in this manual to check out if the meter is defective or not Once defective or malfunctioning is confirmed please contact your local distributor or the...

Отзывы: