![MBT Vaca Shape Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/mbt/vaca-shape/vaca-shape_user-manual_1752812011.webp)
9//22
All rights reserved
Picture
5.2-2
In Picture 5.2-2:
1) A area
:
If handpiece 1 is connected, touch it;
2) B area:
shows working mechanism of handpiece 1;
3) C area:
shows the defaulted parameters of handpiece 1:
i. To adjust cavitation power:
Touch
and it turns to
; Press “
” or “
” to adjust energy.
The energy scope is 0-60W.
You can adjust it on the handpiece too.
ii. To adjust vacuum energy level:
Touch
and it turns to
; Press “
” or “
” to adjust it.
You can adjust it on the handpiece too.
iii. To adjust pulse delay:
Touch
and it turns to
; Press “
” or “
” to adjust it.
You can adjust it on the handpiece too.
iv.
shows temperature of handpiece 1. When the equipment is ready, the indicator light of
handpiece 1 will shine.
Temperature < 50
℃
, the light shows yellow;
Temperature >50
℃
, the light shows red;
Temperature >60
℃
, the handpiece stops working;
4) E area
Press
to change to
situation
Содержание Vaca Shape
Страница 1: ...1 User Manual...