![MBRAUN MB 120B-G Скачать руководство пользователя страница 23](http://html1.mh-extra.com/html/mbraun/mb-120b-g/mb-120b-g_operating-instructions-manual_1752775023.webp)
Connecting the
exhaust
piping for used-up regeneration gas
1. Make
an exhaust
piping between the “Regeneration gas OUTLET” system connection and the factory’s
disposal facility (exhaust).
2. Connection is depressurized.
Caution!
Annoyance by bad smell is expected, as soon as any used-up regeneration gas is escaping to the
surroundings. Environmental pollution and effects detrimental to health, however, are not known,
but cannot be excluded. The manufacturer does not assume any liability!
When using toxic or radioactive material, there should be no discharge of the gas to the surroun-
dings at all!
Connect the
exhaust
piping for vacuum waste gases
1. Make
an exhaust
piping between the vacuum pump exhaust and the disposal facility of the factory (exhaust).
Observe the vacuum pump manufacturer’s instructions for connections.
2. Connection is depressurized.
Note!
Depending on the place where the vacuum pump is used an oil mist filter can be used
instead of the disposal piping. Pertinent information and supplying company may be obtained
from:service@mbraun.
com
Connect the cooling water
Not required in systems with compressor cooling!
1. Make a supply line between the cooling water source and the “Cooling water INLET” system connection by
means of a supply pipe.
Observe chapter 3 “Preparations for Connections”.
2. Make a supply line between the “Cooling water OUTLET” system connection and the depressurized disposal
of water.
3. Turn on the cooling water. The cooling water flow rate to be set depends on the available water temperature,
see chapter 3 “Preparations for connections”.
Connect the system to power
Insert the power plug into the grounded, fuse-protected socket. The required values for connecting should be
taken from the type plate.
!
!
Chapter 4
Installation
24
Содержание MB 120B-G
Страница 6: ...OPERATING INSTRUCTIONS Chapter 1 General Information MBRAUN 1...
Страница 10: ...OPERATING INSTRUCTIONS Chapter 2 Transport Site Selection Modification MBRAUN 2...
Страница 13: ......
Страница 14: ...OPERATING INSTRUCTIONS Chapter 3 Preparing the Connections MBRAUN 3...
Страница 20: ...OPERATING INSTRUCTIONS Chapter 4 Installation MBRAUN 4...
Страница 24: ...OPERATING INSTRUCTIONS Chapter 5 Activating and Deactivating the System MBRAUN 5...
Страница 28: ...OPERATING INSTRUCTIONS Chapter 6 OP 17 Operation Panel MBRAUN 6...
Страница 32: ...OPERATING INSTRUCTIONS Chapter 7 Purging the System MBRAUN 7...
Страница 36: ...OPERATING INSTRUCTIONS Chapter 8 Pressure Control System MBRAUN 8...
Страница 40: ...OPERATING INSTRUCTIONS Chapter 9 Circulation Mode MBRAUN 9...
Страница 45: ......
Страница 46: ...OPERATING INSTRUCTIONS Chapter 10 Regeneration MBRAUN 10...
Страница 51: ......
Страница 52: ...OPERATING INSTRUCTIONS Chapter 11 Settings and Display Patterns MBRAUN 11...
Страница 59: ......
Страница 60: ...OPERATING INSTRUCTIONS Chapter 12 Antechamber Operation MBRAUN 12...
Страница 73: ......
Страница 74: ...OPERATING INSTRUCTIONS Chapter 13 Dust Filters MBRAUN 13...
Страница 77: ......
Страница 78: ...OPERATING INSTRUCTIONS Chapter 14 Analyzers MBRAUN 14...
Страница 90: ...OPERATING INSTRUCTIONS Chapter 15 Maintenance and Service MBRAUN 15...
Страница 94: ...OPERATING INSTRUCTIONS Chapter 16 Spare Parts List MBRAUN 16...
Страница 103: ......
Страница 105: ......