ACV
Range
Precision
Resolution
600 mV
±(0,6%+3)
100 μV
6 V
±(0,6% +3)
1 mV
60 V
10 mV
600 V
±(0,6% +3)
0,1 V
600 V
±(1,0% +5)
1 V
DCA
Range
Precision
Resolu-
tion
600 μA
±(0,12%
+3)
000,1 μA
6000 μA
1 μA
Maximum
current
10 A (For not longer than
15 seconds)
Overload
protection
10 A / 1000 V fuse
"F" marked 10 A
measurement boundary
600mA/1000 V fuse
"F" marked 600 mA-s
measurement boundary
ACA
Range
Precision
Resolution
60 mA
±(0,15%+5)
0,01 mA
600 mA
0,1 mA
Resistance
Range
Precision
Resolution
600 Ω
±(0,7%+5)
0.1 Ω
6 KΩ
±(0,6%+2)
±(0,6%+2)
±(0,7%+2)
±(1,5%+5)
0,001 KΩ
60 KΩ
0,01 KΩ
600 KΩ
0,1 KΩ
6 MΩ
0,001 MΩ
60 MΩ
±(2,5%+6)
0,01 MΩ
Important note:
• In the range of 600 Ω, close the
two instrument lines without any
resistance to be measured, read
the result from the display, and
then subtract the value from the
value displayed by the instrument
to be measured. (This is needed
because the instrument strings
also have resistance and you get
the most accurate value after the
above subtraction). Failure to
display the value for 10 seconds
is normal when the measured
resistance is greater than 1MΩ.
Please wait this is normal for large
MΩ components.