1.0 PRODUCT OVERVIEW

1.1 Description

Maxtec DFB flowmeters are designed and calibrated to provide increased accuracy compared 
to standard flowmeters, when used on air/oxygen blender outlets. By utilizing dual-scale 
graduations, the flowmeters provide two flowmeters in one for increased accuracy at lower 
flows. The flowmeters feature high-quality acrylic block bodies and precision valves. Each 
flowmeter comes equipped with a DISS outlet and DISS adapter for neat and compact attach-
ment to the side of the blender.

1.2 Intended use

The flowmeter is intended for use as a secondary flowmeter for an air/oxygen blender, such 
as the Maxtec MaxBlend 2, and is to be used by physicians, respiratory therapists and other 
authorized hospital personnel to administer selected doses of medical gases to a patient.

2.0 SPECIFICATIONS

2.1 Flow Range Accuracy Chart 

MODEL

FLOW

GRADUATIONS

ACCURACY

R219P99-400

0-3 LPM

0.1 (0.1-1) LPM

0.5 (1-3) LPM

± 0.5 LPM

R219P79-400

0-15 LPM

0.25 (0.5-3) LPM

1 (5-15) LPM

0.5-3: ± 0.5 LPM

5-15: ± 10% of indicated value

R219P88-400

0-30 LPM

0.25 (0.5-3) LPM

2.5 (5-30) LPM

0.5-3: ± 0.5 LPM

5-30: ± 10% of indicated value

R219P87-400

0-70 LPM

1 (2-15) LPM

5 (15-70) LPM

2-4: ± 0.5 LPM

5-70: ± 10% of indicated value

2.2 Transport/Storage Requirements 

-40 °F (-40 °C) to 140 °F (60 °C)

NOTE:

 Storage/transport outside the specified range may cause damage to the flowmeter.

The above flowmeter models are calibrated for air/oxygen blenders at 70°F (21°C) and stan-
dard atmospheric pressure with inlet pressures at 50 PSI and 60% O₂. Specifications are subject 
to change without prior notice.

3.0 OPERATING INSTRUCTIONS

3.1 General Operation

1.  Remove the air and oxygen line pressure to the air/oxygen blender and ensure 

the valve to the flowmeter is closed.

2.  Apply a thread sealant that is appropriate for medical oxygen use to the male NPT 

thread of the supplied 90 degree NPT to DISS adapter.

3.  Use a backing-wrench to secure the female NPT fitting on the back of the 

flowmeter while hand-tightening the male NPT thread of the 90 degree NPT to 
DISS adapter into the female NPT fitting.

4.  With the backing-wrench still in place, tighten the 90 degree NPT to DISS adapter 

fitting at least one full revolution. Continue to tighten the fitting until it is aligned 
properly for the desired attachment configuration on the air/oxygen blender 
output.

5.  Attach the flowmeter to the air/oxygen blender output using the DISS connection. 

The flowmeter must be mounted vertically for accurate measurements.

6.  Restore pressure to the air/oxygen blender and verify that the float ball is at the 

very bottom of the flow tube.

7.  NOTE: If the float is not resting at the bottom of the flow tube, the product is 

leaking; consult the “TROUBLESHOOTING” Section 4.0.

8.  Adjust flow:  

  To increase- Turn knob counterclockwise 
  To decrease- Turn knob clockwise

9.  Set the flow by aligning the center of the float ball with the indicator lines on the 

flow tube. Adjusting flow beyond the last calibrated indicator line will result in an 
undetermined flow.

4.0 TROUBLESHOOTING

4.1 Troubleshooting Table

PROBLEM

CAUSE

REMEDY

Will not shut off

Leak
Defective valve

Consult your  

provider or Maxtec

Sticking float ball or unable 
to set desired flow

Debris in flow tube

Knob will not turn

Valve seized

5.0  RECOMMENDED MAINTENANCE

5.1 Cleaning Instructions

 

Depressurize and disconnect all connections before cleaning.

 

Clean exterior surfaces of the flowmeter with a cloth dampened with a mild 
detergent and water.

 

Wipe dry with a clean cloth.

 

CAUTION:

 Do Not Autoclave

Содержание R219P79-400

Страница 1: ...eters D E S I G N E D F O R B L E N D E R S D U A L TA P E R EN Instructions for Use 2 FR Instructions d utilisation 4 DE Bedienungsanleitung 6 NL Gebruiksaanwijzingen 8 IT Istruzioni per l uso 10 ES...

Страница 2: ...dthisUserManualbeforeinstallingoroperatingtheflowmeter To Reduce the Risk of Fire or Explosion ALWAYS follow ANSI and CGA standards for Medical Gas Products and flowmeters and OxygenHandling DO NOT us...

Страница 3: ...C and stan dardatmosphericpressurewithinletpressuresat50PSIand60 O Specificationsaresubject to change without prior notice 3 0 OPERATING INSTRUCTIONS 3 1 General Operation 1 Remove the air and oxygen...

Страница 4: ...neIRM LisezceManueld utilisationavantd installeroud utiliserled bitm tre Afin de r duire les risques d incendie ou d explosion TOUJOURSrespecterlesnormesdel ANSIetdelaCGAsurlesproduitsetd bitm tres ga...

Страница 5: ...ressionatmosph riquestandardavecdespressionsenentr ede3 4bar 50 psi et 60 d O2 Ces caract ristiques techniques peuvent tre modifi es sans pr avis 3 0 INSTRUCTIONS D UTILISATION 3 1 Utilisation en cond...

Страница 6: ...sk of Fire or Explosion Befolgen Sie STETS die ANSI und CGA Normen f r medizinische Gasprodukte Durchflussmesserundf rdenUmgangmitSauerstoff KEIN L Schmierfett organische Schmierstoffe oder brennbare...

Страница 7: ...h nischen Daten k nnen ohne vorherige Benachrichtigung ge ndert werden 3 0 BETRIEBSANWEISUNG 3 1 Allgemeiner Betrieb 1 Beseitigen Sie den Druck in der Luft und Sauerstoffleitung zum Luft Sauerstoffmis...

Страница 8: ...de resultaten van eenMRIkanbe nvloeden Leesdezegebruikershandleidingvoordatudedebietmeterinstalleertofgebruikt Om het risico op brand of explosie te verminderen Volg ALTIJD de ANSI en CGA normen voor...

Страница 9: ...aatdrukvan50psien60 O2 Specificaties kunnen zonder voorafgaande kennisgeving worden gewijzigd 3 0 GEBRUIKSAANWIJZING 3 1 Algemene bediening 1 Verwijder de lucht en zuurstoflijndruk naar de lucht zuurs...

Страница 10: ...ltati di unaRM Leggereilpresentemanualeperl usoprimadiinstallareomettereinfunzioneilflussometro Per ridurre il rischio di incendio o esplosione SeguireSEMPREglistandardANSIeCGAperigasditipomedicale if...

Страница 11: ...ggette a modifiche senza preavviso 3 0 ISTRUZIONI OPERATIVE 3 1 Funzionamento generale 1 Rimuovere la pressione alla linea di aria e ossigeno al miscelatore aria ossigeno e controllare che la valvola...

Страница 12: ...talaci nodelfuncionamientodelcaudal metro Para reducir el riesgo de fuego o de explosi n Siga SIEMPRE las normativas ANSI y CGA para Productos de gases m dicos y caudal metros ymanipulaci ndelox geno...

Страница 13: ...radade3 4bar 50 psi y 60 de O2 Las especificaciones est n sujetas a cambio sin previo aviso 3 0 INSTRUCCIONES DE FUNCIONAMIENTO 3 1 Funcionamiento general 1 Retire la presi n de la l nea de aire y ox...

Страница 14: ...14 This page intentionally left blank...

Страница 15: ...15 This page intentionally left blank...

Страница 16: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Отзывы: