Maxtec MaxO2ME Скачать руководство пользователя страница 9

WWW.MAXTEC.COM  •  (800) 748-5355 

9

 ENGLISH

3.4 Backlight Operation

To turn on the backlighting:

1.  When the unit is on, pressing the Backlight 

 key will turn the backlighting on for 

30 seconds. Additional presses will turn off the backlighting.

2.  If the device is being used in a dark location, press any key to activate the back light.

 CAUTION:

 Excessive use of the backlight can reduce the life of the batteries.

3.5 Sleep Mode Operation

To use the sleep mode function:

1.  Remove the batteries from the unit.
2.  Locate the sleep mode switch in the battery compartment and set to the ON position.
3.  Replace the batteries in the unit.

The unit will now perform a normal boot-up operation with sleep mode enabled. With sleep 
mode enabled the unit will function with all the same parameters as outlined above with one 
new feature. While in the ON mode, the unit will time-out after 90 seconds to a battery saving 
condition. This condition will be indicated by a crescent moon on the display. While in this 
condition any key that is pressed will return the unit to the ON mode and reset the 90 second 
time-out counter. In sleep mode, the device will continue to monitor the oxygen level and will 
activate the alarm if an alarm condition occurs.

3.6 External Power Supply Operation

To extend the life of the batteries a Maxtec approved external power supply can be purchased. 
Once connected to the unit, total power is supplied by the external power supply. The batter-
ies are still required to be in the unit and will provide emergency power in the event main AC 
power is lost. 

NOTE:

 Use only the Maxtec approved external power supply in Section 9.0 Spare Parts and 

Accessories.

NOTE:

 The power supply is not a battery charger.   

DO NOT 

use rechargeable batteries.

WARNING:

 Do not position the equipment in a way that it would be difficult to unplug the 

power supply.  Unplugging the power supply is the only means of disconnecting or isolating 
the equipment from AC mains power.

4.0 SENSOR REMOVAL AND 

REPLACEMENT

The MaxO2 ME is shipped with a new Max 550E oxygen sensor. 

Although the sensor has a very long expected life, eventually the sensor will require replace-
ment. Removing or installing a sensor, when necessary, is a very simple procedure. 

To remove and install a new sensor:

1.  Grasp the sensor in one hand and, with the other hand, unscrew the cable connector 

counter-clockwise at the sensor.

2.  Pull out the cable connector plug from the expired sensor.
3.  Unscrew the flow diverter from the sensor and discard the expired sensor or return it 

to Maxtec for proper disposal.

NOTE:

 The sensor contains lead and lead acetate, be sure to dispose of expired sensors in 

accordance with hospital, local, state and federal regulations.

4.  Remove the new sensor from the packaging and remove the protective film from the 

sensor face.

5.  Insert the cable connector plug into the receptacle of the new sensor and tighten the 

cable connector.

6.  Screw the flow diverter onto the new sensor.
7.  Wait approximately 20 minutes for the sensor to reach equilibrium.
8.  Calibrate the new sensor.

NOTE:

 If the monitor is on when the sensor is detached and replaced, the monitor will auto-

matically force a re-calibration. The display will read “CAL”.

NOTE:

 If the cable locking nut is not fully fastened onto the sensor, then the sensor may not 

function properly.

5.0 PROBLEM SOLVING

The MaxO2 ME monitors have a self test feature built into the software to detect faulty calibra-
tions, oxygen sensor failures, and low operating voltage. These are listed below, and include 
possible actions to take, if an error code occurs.

NOTE:

 The operator must be facing the device and positioned within 4 meters to distinguish 

the visual alarm indicators. Audible alarms can be distinguished as long as the operator is in 
the same room and the ambient noise level is typical for a clinical setting.

LOW BATTERY ICON 

:

 If the low battery icon is displayed on the LCD readout at any time, 

the batteries should be replaced as quickly as possible.

E01:

 Calibration error, sensor output lower than expected. See note below.

E02:

 No sensor attached. Reconnect sensor, see note below.

E03:

 No Valid Calibration Data Available, make sure unit has reached thermal equilibrium and 

perform a calibration routine.

E04:

 Battery Below Minimum Operating Voltage, replace batteries. A medium priority alarm will 

sound every 25 seconds until the batteries are replaced or become too dead to sound the alarm.

E05:

 Calibration error, sensor output higher than expected. See note below.

E06:

 Non-compatible oxygen sensor. Reconnect sensor, see note below.

E07:

 Calibration error, sensor output is not stable. See note below.

E08:

 Calibration error, battery too low to preform calibration. Replace batteries and 

re-calibrate.

NOTE:

 If you receive a E01, E05, or an E07 error code, correct by ensuring the calibration gas is 

either room air or 100% oxygen. Also ensure the calibration gas flow, pressure and concentra-
tion is constant. Allow sufficient time for the sensor to stabilize in the calibration gas and with 
room temperature, then attempt to calibrate again.

If these steps do not correct the error, contact Maxtec for technical support.

NOTE:

 Use only a Maxtec approved Max-550E sensor called out in Section 9.0 Spare Parts and 

Accessories. The Max 550E sensor is equipped with an authentication chip to ensure the moni-
tor is used with an approved sensor.

NOTE:

 Correcting E02 or E06 errors:

1.  Disconnect the sensor and reconnect, making sure the male plug is fully inserted into 

the receptacle before tightening the threaded locking shroud. The analyzer should now 
perform a new calibration with the error cleared.

2.  If the error still persists, remove the batteries and external power, wait 30 seconds, 

then reinstall to perform a factory reset and diagnostic on the analyzer. The analyzer 
should again perform a new calibration with the error cleared.

3.  Contact Maxtec Customer Service Department if the error code cannot be cleared.

Содержание MaxO2ME

Страница 1: ...MaxO2 ME Instructions for Use E N G L I S H R230M01 001 REV H...

Страница 2: ...60601 1 2formedicalproducts Theselimitsprovidereasonableprotectionagainstelectromag netic interference when operated in the intended use environments described in this manual MRI Notice Thisequipment...

Страница 3: ...comes in con tact with infectious agents Gasleaksthatcauseroomairtomixwiththegassamplemaycauseinaccurateoxygenread ings EnsuretheO ringsonthesensorandflowdiverterareinplaceandintactpriortouse DONOTexp...

Страница 4: ...g Oxygen Calibration 7 3 0 OPERATING INSTRUCTIONS 7 3 1 Alarm Setting Procedure 7 3 1 1 Low Alarm Setting 7 3 1 2 High Alarm Setting 8 3 1 3 Smart Alarm Mode 8 3 2 Basic Operation 8 3 3 Alarm Conditio...

Страница 5: ...mponent Identification 1 LOWALARMLED Inalowalarmcondition theyellow LOWALARM LEDwillflashonce every two seconds accompanied by the audio buzzer If the Oxygen level is below 18 the red LOW ALARM LED wi...

Страница 6: ...sensor after calibration may shift the calibra tion point enough to require recalibration CAUTION The flow diverter for the sensor is for use with flowing gases only DO NOT use the diverter when perf...

Страница 7: ...1 Allow adequate time for the sensor to equilibrate to a new ambient temperature 2 When used in a breathing circuit place the sensor upstream of the heater 3 For best results perform the calibration...

Страница 8: ...from the low alarm setting Pressing the Down key will subtract one from the high alarm setting and add one to the low alarm setting In other words the Up Arrow widens the alarm band and the down arro...

Страница 9: ...the new sensor and tighten the cable connector 6 Screw the flow diverter onto the new sensor 7 Wait approximately 20 minutes for the sensor to reach equilibrium 8 Calibrate the new sensor NOTE If the...

Страница 10: ...axtec Certified Service Technician 6 3 Replacing Sensor Cable After extended use or abuse to the sensor cable the cable may begin to wear and lose its ability to properly retract The cable can be remo...

Страница 11: ...or the reading to stabilize before calibration Also note that the monitor may sense that the concentration is changing and an E07 error code will display Sensors come supplied with a flow diverter The...

Страница 12: ...AC mains input lines IEC 61000 4 11 Dip 95 0 5 periods Dip 60 5 periods Dip 30 25 periods Dip 95 5 seconds Recommended separation distances between portable and mobile RF communications equipment and...

Страница 13: ...os amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic si...

Страница 14: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Отзывы: