Maxtec MaxO2ME Скачать руководство пользователя страница 3

WWW.MAXTEC.COM  •  (800) 748-5355 

3

 ENGLISH

 

Never autoclave, immerse in liquid or expose the MaxO2 ME (including sensor) to high tem-

peratures (>50°C). Never expose the device to liquid, pressure, irradiation vacuum, steam, 

or chemicals.

 

To protect the unit from potential leaky battery damage always remove batteries when the 

unit is going to be stored (not in use for 30 days or more) and replace dead batteries with 

recognized name brand AA Alkaline batteries.

 DO NOT 

use rechargeable batteries.

 DO NOT  

attempt to replace oxygen sensor or batteries while the device is in use.

 

This device does not contain automatic barometric pressure compensation.

 

Not for use in an MRI environment.

 

Battery replacement by inadequately trained personnel could result in a safety hazard.

 

Electrical shock or damage to the equipment may occur if an inappropriate external power 

supply is used. Maxtec recommends using only the Maxtec approved external power supply, 

as listed in 9.0 Spare Parts and Accessories.

NOTE:

 The MaxO2 ME oxygen monitor has been manufactured with a low alarm setting adjust-

able down to 15% that requires deliberate action to set it below 18%. See section 3.1 Alarm 
Setting Procedure.

 DO NOT 

clean or dry the MaxO2 ME with a high pressure air gun. Applying high pressure 

air to the MaxO2 ME may damage components and render the system inoperable.

 DO NOT 

over clean the MaxO2 ME. Repeated use of a cleaning agent can cause residue 

buildup on critical components. Excessive residue buildup can affect the MaxO2  ME's 

performance.

 

When cleaning the MaxO2 ME:   

DO NOT

 use harsh abrasives.   

DO NOT

 immerse the 

MaxO2 ME in liquid sterilizing agents or liquids of any kind.   

DO NOT

 spray cleaning solu-

tion directly onto the device.   

DO NOT

 allow cleaning solution to pool on the device.

 DO NOT 

sterilize the MaxO2  ME. Standard sterilization techniques may damage the 

monitor.

 

If the MaxO2 ME does not function as outlined in section 2.0, contact a Maxtec trained service 

technician or Maxtec for service.

  DO NOT 

allow the sensor to come in contact with exhaled patient gases or other potential 

sources of contamination. The sensor face cannot be decontaminated if it comes in con-

tact with infectious agents.

 

Gas leaks that cause room air to mix with the gas sample may cause inaccurate oxygen read-

ings. Ensure the O-rings on the sensor and flow diverter are in place and intact prior to use.

  DO NOT 

expose the sensor face to liquids or allow humidity to condense on the face of the 

sensor as this may impair the function of the MaxO2 ME.

 

The MaxO2 ME and sensor are non-sterile devices.

 

Regularly inspect the MaxO2 ME and associated components for damage or electrolyte leak-

age prior to use. 

 DO NOT 

use if damaged.

 DO NOT 

obstruct alarm.

 DO NOT 

smoke in an area where oxygen is being administered.

 

The MaxO2 ME may only be calibrated using 20.9% oxygen (room air) or 100% oxygen. 

Calibration at other concentrations will result in inaccurate readings.

 

The oxygen sensor should be operated in an upright position (sensor face downwards). 

Operating the oxygen sensor upside down may cause the sensor to function improperly.

 

When using the approved external power supply, functional batteries must also be installed 

in the device. The device will not operate solely on the external power supply.

 

In the event of exposure to an ELECTROMAGNETIC DISTURBANCE the analyzer may display an E06 

or E02 error message. If this occurs, refer to Section 5.0 for instructions to resolve the problem.

SYMBOL GUIDE

The following symbols and safety labels are found on the MaxO2 ME:

Attention, consult 
accompanying documents

On/Off Key

Consult Instructions For Use

Calibration Key

Do Not

Calibration Reminder

Unlock Key

Silent Key

Backlight Key

Smart Alarm Key

High Alarm Indicator

Low Alarm Indicator

Smart Alarm Mode Indicator

Alarm Silence Indicator

Down (Low Alarm) Key

Up (High Alarm) Key

Low Battery Indicator

Sleep Mode Indicator

Federal law (USA) restricts 
this device to sale by or 
on order of a physician

Meets ETL Standards

Manufacturer

Direct Current

Date of Manufacture

Medical Device

Warning

Two means of patient 
protection (double insulated)

Caution

Power Supply Meets CEC Tier 
3 and EU Phase 2 Standards

Conforms to EU Requirements

Authorized 
Representative in the 
European Community

For use in dry indoor locations

50°C

(122°F)

5°C

(41°F)

Storage Temperature 
Range

Serial Number

Catalog Number

Complies with Directive 
2011/65/EU

Ingress Protection Rating

Combined UL / CSA Mark

Corrosive

Do not throw away. Follow 
local guidlines for disposal.

Type B Applied Parts

MR Unsafe

Содержание MaxO2ME

Страница 1: ...MaxO2 ME Instructions for Use E N G L I S H R230M01 001 REV H...

Страница 2: ...60601 1 2formedicalproducts Theselimitsprovidereasonableprotectionagainstelectromag netic interference when operated in the intended use environments described in this manual MRI Notice Thisequipment...

Страница 3: ...comes in con tact with infectious agents Gasleaksthatcauseroomairtomixwiththegassamplemaycauseinaccurateoxygenread ings EnsuretheO ringsonthesensorandflowdiverterareinplaceandintactpriortouse DONOTexp...

Страница 4: ...g Oxygen Calibration 7 3 0 OPERATING INSTRUCTIONS 7 3 1 Alarm Setting Procedure 7 3 1 1 Low Alarm Setting 7 3 1 2 High Alarm Setting 8 3 1 3 Smart Alarm Mode 8 3 2 Basic Operation 8 3 3 Alarm Conditio...

Страница 5: ...mponent Identification 1 LOWALARMLED Inalowalarmcondition theyellow LOWALARM LEDwillflashonce every two seconds accompanied by the audio buzzer If the Oxygen level is below 18 the red LOW ALARM LED wi...

Страница 6: ...sensor after calibration may shift the calibra tion point enough to require recalibration CAUTION The flow diverter for the sensor is for use with flowing gases only DO NOT use the diverter when perf...

Страница 7: ...1 Allow adequate time for the sensor to equilibrate to a new ambient temperature 2 When used in a breathing circuit place the sensor upstream of the heater 3 For best results perform the calibration...

Страница 8: ...from the low alarm setting Pressing the Down key will subtract one from the high alarm setting and add one to the low alarm setting In other words the Up Arrow widens the alarm band and the down arro...

Страница 9: ...the new sensor and tighten the cable connector 6 Screw the flow diverter onto the new sensor 7 Wait approximately 20 minutes for the sensor to reach equilibrium 8 Calibrate the new sensor NOTE If the...

Страница 10: ...axtec Certified Service Technician 6 3 Replacing Sensor Cable After extended use or abuse to the sensor cable the cable may begin to wear and lose its ability to properly retract The cable can be remo...

Страница 11: ...or the reading to stabilize before calibration Also note that the monitor may sense that the concentration is changing and an E07 error code will display Sensors come supplied with a flow diverter The...

Страница 12: ...AC mains input lines IEC 61000 4 11 Dip 95 0 5 periods Dip 60 5 periods Dip 30 25 periods Dip 95 5 seconds Recommended separation distances between portable and mobile RF communications equipment and...

Страница 13: ...os amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic si...

Страница 14: ...2305 South 1070 West Salt Lake City Utah 84119 800 748 5355 www maxtec com...

Отзывы: