MTS600 Instruction Manual
15
the reticle. The last adjustment of each fine motion screw should be in the
clockwise direction.
④
Readjust the focus until there is no parallax: readjust the focus with the
focusing ring until there is no parallax between the target image and the
reticle.
Caution
:
l
When sighting the target, strong light shining directly into the objective
lens may cause the instrument to malfunction. Protect the objective lens
from direct light by attaching the lens hood.
3.4 Power ON/OFF
①
Confirm that the instrument is leveled and centered precisely.
②
Press the power key, instrument is powered on. The instrument will make a
sound of beep, a self-check is run ,while the screen display the instrument
model information, instrument number and the software version number. if
the instrument is normal, screen display "Turn telescope" prompt.
③
Rotate telescope, the instrument will make a sound of beep, and vertical
angle index setting finished, enter basic measurement mode.
Caution:
l
When the battery capacity is insufficient to support the instrument
normal working, press the power key, the screen will display
“
battery is
low
”
, and the instrument will turn power off automatically.
l
After power on, you should pay attention to the mark indicates the
battery power remaining on screen. Confirm that the battery power is
enough for work, otherwise you should replace the battery or charge the
battery.
l
If
“
over range
”
is displayed, the instrument tilt sensor is indicating that
the instrument is out of level, need to be leveled once again.
l
Due to vibration or strong wind, the angle display is unsteady. You
should turn off the tilt angle compensation before measurement. please
see below for details.