- 31 -

 

 

5-3. Change of the measuring mode (only for fumigation 

gases) 

The measuring mode of the detector (for fumigation gases) can be switched from "concentration measuring 
mode" to "differential measuring mode", where lower concentration detection can be conducted. 
 

On the "MODE CHG" 
screen in the display mode, 
press the POWER button. 

The detector enters the setting 
to switch the measuring mode. 

 

 

 

 

 

 

 

Press the MODE or AIR 
button to select the 
measuring mode. 

The measuring mode switches 
to the other every time the 
button is pressed. 

 

 

 

 

<Concentration measuring 

mode> 

<Differential 

measuring mode> 

 

 

 

 

In the appropriate 
measuring mode, press the 

POWER button. 

"END" is displayed, returning to 
the display mode. 

 

 

 

 

Содержание 8058 FUM

Страница 1: ...INT 0321 XX REV A 01 28 16 Portable Gas Leak Detector 8058 Series Operating Manual ...

Страница 2: ...the alarm and operation sounds 22 4 12 How to turn on the illumination lamp 23 4 13 Power off 23 5 How to Set Display Mode 24 5 1 Entering the display mode 24 5 2 Concentration displayed gas reading setting 27 5 3 Change of the measuring mode only for fumigation gases 31 5 4 Log data display 32 6 How to Set User Mode 33 6 1 Entering the user mode 33 6 2 Date time setting 35 7 Alarm function 36 7 1...

Страница 3: ...etector is used to detect a single gas such as phosphine PH3 etc in air It provides two different specifications TYPE SC used to detect leaking of general gases including semiconductor material gases and TYPE FUM used to detect leakage of fumigation gases Detection results are not intended to guarantee life or safety in any way Definition of DANGER WARNING CAUTION and NOTE Throughout this manual t...

Страница 4: ...t in a manhole or confined space do not lean over or look into the manhole or closed space It may lead to dangers because oxygen deficient air or other gases may blow out Oxygen deficient air or other gases may be discharged from the gas exhausting outlet of the detector Never inhale the air or gases High concentration 100 LEL gases may be discharged from the gas exhausting outlet of the detector ...

Страница 5: ...ith a temperature of 20ºC and a humidity of 40 RH and taken outside a temperature of 30ºC and a humidity of 60 RH an alarm may be triggered by humidity changes In this case perform the air calibration outside and measure The detector performs an air calibration automatically when it is turned on Even in this case check the atmosphere for conditions and perform air calibration again in the measured...

Страница 6: ...f the detector in a place where it is exposed to direct sunlight Do not store the detector in a sun heated car Avoid a sudden humidity or temperature change Do not leave the detector in a place with high temperature and humidity for a long time The performance of the detector may deteriorate Observe the operating restrictions to prevent condensation inside the concentration meter or gas sampling h...

Страница 7: ...ng it again If the above 4 gas high concentration solvent gas is exposed a crack may occur in the detector Thus do not use high concentration solvent gases About battery replacement Always turn off the power before replacing the batteries Replace both of the batteries with new ones at one time Pay attention to the polarities of the batteries Usage In a low temperature environment the operating tim...

Страница 8: ...ase numerical values in the setting mode 5 MODE button Press this button to change between display modes Used to change the screen and increase numerical values in the setting mode 6 POWER button Turns the power ON or OFF Used to change the screen and confirm values in the setting mode 7 Gas outlet Exhausts the gas drawn into the detector Do not block it 8 Alarm LED arrays infrared port Blinks or ...

Страница 9: ...ached AA alkaline battery 2 pcs installed Taper nozzle 1pc Hand strap 1pc Water proofing filter for exchange 5pcs Operating manual English 1pc Operating manual Japanese 1pc Product warranty 1pc If there is anything missing contact Matheson CAUTION Do not remove the panel sheet on the display The water proof and dust proof performances will be deteriorated Do not affix a label or the like on the in...

Страница 10: ...display Displays the setting status of the operation sound 6 Battery level icon Displays a reference of the battery level 7 Bar meter display Displays the level of gas concentration with the bar meter 8 Mode display Displays the measuring mode status 9 Gas name and message display Displays the gas name and a message according to the function NOTE The meanings of battery level icons are as follows ...

Страница 11: ...ted Check that the taper nozzle is not bent or has no hole 4 3 How to replace the batteries When the detector is used for the first time or when the battery level is low install two new AA alkaline batteries according to the following procedures 1 Check that the power of the detector is turned off Turn off the power if it is turned on 2 Turn the lock plate counterclockwise and open the battery cov...

Страница 12: ...th new ones at one time Pay attention to the polarities during replacement Replace while checking the battery polarities stamped on the body If the battery cover is not completely locked the dry batteries may drop off or water may get in through the clearance Water may also get in if a minute foreign substance is caught between the detector and battery cover ...

Страница 13: ...n below and the detector enters the detection mode Display example For Phospine 1 Press and hold the POWER button for one second or longer Hold it down until all the LCDs and alarm lamp light up and the buzzer blips All LCDs light up Date time display Calibration gas name display Full scale Display Display example Thursday October 01 2015 12 34 Display example PH3 Display for fumigation gases Dete...

Страница 14: ...ty status of the sensor If the detector is not used for a long period warm up time may be longer or a sensor abnormality FAIL SENSOR may be displayed In such a case restart the detector The detector performs an air calibration automatically on start up If an environment where the detector is turned on is different from the measured atmosphere perform air calibration again in the measured atmospher...

Страница 15: ... 15 4 5 Basic operating procedures The detection mode is used after power on Display example For Phosphine ...

Страница 16: ... pressure is applied to the gas inlet and outlet of the detector detected gases may leak out from its inside and may cause dangerous conditions Be sure that excessive pressure is not applied to the detector while used When air calibration is performed in the atmosphere check the atmosphere for contaminates before beginning it If interference gases exist the adjustment cannot be performed properly ...

Страница 17: ...ue showing the correlation between a leakage amount and the bar display If the measured atmosphere changes perform air calibration again Differential measuring mode Perform automatic air calibration at regular intervals in the atmosphere of measurement environment This is a high sensitivity mode that enables detection of ultralow amount of leaking gas in the air by eliminating the environmental ef...

Страница 18: ...ons at the same time The alarm set point changes by two every time the AIR and MODE buttons are pressed Gas alarm set point for differential measuring mode only for fumigation gases 1 In the detection mode press the AIR and MODE buttons at the same time The alarm set point changes in six steps every time the AIR and MODE buttons are pressed ...

Страница 19: ...on page 31 CAUTION Keeping contact with a gas with the same concentration will decrease the bar meter by the automatic air calibration function For excessive temperature humidity changes the bar meter may increase When the mode is switched from the differential mode to the concentration measuring mode check the atmosphere for cleanliness and perform air calibration again in the measured atmosphere...

Страница 20: ... is not contaminated Display example For fumigation gases 1 In the detection mode hold down the AIR button Display for fumigation gases 2 Release the AIR button when the display changes from HOLD AIR to RELEASE Buzzer sound Three times blip blip blip The air calibration is done and the gas detector returns to detection mode Buzzer sound Once blip If air calibration fails FAIL AIR CAL is displayed ...

Страница 21: ...to 256 points of data can be recorded When the number of recorded data points reaches the maximum recorded data will be overwritten starting from the oldest data 1 In the detection mode press the MODE and POWER buttons at the same time The detector enters the station ID selection screen 2 Use the MODE or AIR button to select the station ID and press the POWER button The peak value is displayed To ...

Страница 22: ... longer NOTE To clear a retained peak value hold down the MODE button in the detection mode for one second 4 11 Change of the alarm and operation sounds Turn the alarm and operation sounds ON or OFF Display example For fumigation gases 1 In the detection mode press the AIR and POWER buttons at the same time The settings for the alarm and operation sounds are switched every time the AIR and POWER b...

Страница 23: ...tomatically go off in about two minutes after it lights up To turn off the illumination lamp hold down the AIR and POWER buttons at the same time for three seconds or longer 4 13 Power off Press and hold the POWER button at least three seconds until the buzzer blips three times TURN OFF disappears to turn off the power CAUTION If a gas remains in the detector it enters the automatic exhaust mode f...

Страница 24: ...e display mode 2 Next press the MODE button to display the appropriate menu The setting screen for the display mode is switched every time the button is pressed 3 Select the setting item and press the POWER button For setting items see Display mode overview on page 25 NOTE The detector automatically returns to the detection mode in about 20 seconds if the detector is left unoperated Gas detection ...

Страница 25: ...gas reading setting By changing the setting to the pre registered gas in the detector the converted concentration from the detection target gas PH3 will be displayed P 27 Measuring mode setting only for fumigation gases Changes in small measure the pump suction volume by mode exchanging in the differential measuring mode the pump suction volume is up P 31 NORMAL Concentration measuring mode DIFF D...

Страница 26: ... 26 Entering user mode Enters the user mode P 33 Detection mode Returns to the detection mode ...

Страница 27: ...play example For fumigation gases 1 On the GAS LIST screen in the display mode press the POWER button The detector enters the gas reading setting The currently set gas name is displayed 2 Press the MODE or AIR button to select a gas name to be read instead The gas name switches to another every time the button is pressed 3 When the target gas name is displayed press the POWER button END is display...

Страница 28: ...ride or sulfur compound will shorten the sensor life or cause larger errors If a Si compound is detected the sensitivity will decrease shortening the sensor life Even for a gas shown only with the bar meter display but no scales the meter reads Scale 2 or 3 Use it as an indication of concentration increase If a high concentration solvent gas is drawn the rubber seal or other part used in the detec...

Страница 29: ...w NOTE Sulfur dioxide SO2 1 Sulfur compound See the below NOTE Propane C3H8 5 20 R 134a R 134a 50 250 R 22 R 22 10 50 Chloride See the below NOTE R 32 R 32 10 50 Normal hexane n C6H14 10 50 Benzene C6H6 0 5 10 Formaldehyde HCHO HCHO 10 50 Methane CH4 1 20 Methyl alcohol CH3OH 1 10 Methyl ethyl ketone MEK 1 10 Hydrogen sulfide H2S 0 1 Sulfur compound See the below NOTE Diborane B2H6 0 1 Germane GeH...

Страница 30: ...cause larger errors If a Si compound is detected the sensitivity will decrease Even for a gas shown only with the bar meter display but no scales the meter reads Scale 2 Use it as an indication of concentration increase If a high concentration solvent gas is drawn the rubber seal used in the detector will deteriorate ...

Страница 31: ...n be conducted 1 On the MODE CHG screen in the display mode press the POWER button The detector enters the setting to switch the measuring mode 2 Press the MODE or AIR button to select the measuring mode The measuring mode switches to the other every time the button is pressed Concentration measuring mode Differential measuring mode 3 In the appropriate measuring mode press the POWER button END is...

Страница 32: ...ation ID and memory number are displayed alternately 2 Press the MODE or AIR button to select log data to be displayed The recorded content switches to the other every time the button is pressed Station ID Memory number 3 When the target log data is displayed press the POWER button The gas name and peak value of the selected memory are displayed alternately 4 To exit the display press the MODE and...

Страница 33: ...times to display entering user mode press the POWER button The detector enters the date time setting in the user mode 2 Next press the MODE button to display the appropriate menu The setting screen for the user mode switches to another every time the button is pressed 3 Select the setting item and press the POWER button For setting items see User mode overview on page 34 CAUTION Return to the dete...

Страница 34: ...etting Set the date time of the internal clock P 35 ROM SUM display Displays the program number and SUM value of the detector This is not typically used by the user Entering detection mode To exit the user mode press the POWER button to enter the detection mode ...

Страница 35: ...r mode press the POWER button The detector enters the date time setting 2 Press the MODE or AIR button set the date time and press the POWER button 3 Set year month day hour minute in this order When the minute value is confirmed END is displayed and then the detector returns to the user mode menu ...

Страница 36: ... details is displayed on the LCD System abnormalities FAIL SYSTEM Sensor abnormalities FAIL SENSOR Calibration abnormalities FAIL AIR CAL Low battery voltage FAIL BATTERY Low flow rate FAIL LOW FLOW Clock abnormalities FAIL CLOCK Pump abnormalities FAIL PUMP Alarm lamp Repeatedly blinks at about one second intervals Buzzer Repeatedly sounds intermittent blips at about one second intervals Blip bli...

Страница 37: ...or more for every six months to maintain the performance as a safety unit Maintenance item Maintenance content Daily maintenance Monthly maintenance Regular maintenance Battery level check Check that the battery level is sufficient Concentration display check Draw in fresh air and check that the concentration display value is zero When the reading is incorrect perform a zero adjustment by air cali...

Страница 38: ...eck Verifies that the concentration display value is zero by using zero gas Performs the air calibration if the reading is incorrect Flow rate check Checks the flow rate indicator to find abnormalities Checks the flow rate by using an external flow meter to verify the accuracy of the flow rate indicator on the detector If the flow rate is incorrect performs the flow rate adjustment Filter check Ch...

Страница 39: ...hake it about ten times with the buzzer sound opening facing downward 3 Wipe away moisture coming out from the inside thoroughly using a towel cloth etc 4 Place the gas detector on a dry towel cloth etc and let it stand at normal temperatures 8 3 Parts replacement Filter replacement procedure Continuing to use the detector may cause the filter to be contaminated or clogged Replace the filter if it...

Страница 40: ... the main unit Sensor replacement The built in sensors of the detector have a life time and must be replaced regularly within one year The sensor life has expired if for example the sensors cannot be calibrated in span adjustment the readings do not come back after air calibration or the readings fluctuate In this case contact Matheson Battery replacement For battery replacement see How to replace...

Страница 41: ...t must be treated properly as an industrial waste in accordance with the local regulations etc CAUTION When the detector is used again after a long period storage always perform a calibration Contact Matheson for information on readjustment including calibration WARNING Dispose of dry batteries in accordance with procedure specified by the local authority CAUTION If the detector is not used for a ...

Страница 42: ...hat the batteries are properly installed to the main unit Abnormal operations Disturbances by sudden static electricity noise etc Turn off the power once and then turn it on again restart Cannot operate the gas detector Disturbances by sudden static electricity noise etc Remove the batteries in a safe place Then reinstall them and turn on the power to perform operations System abnormalities FAIL S...

Страница 43: ... contact Matheson to replace the pump The pump has deteriorated Contact Matheson to replace the pump Air calibration impossible FAIL AIR CAL Fresh air is not supplied around the detector Supply fresh air and then perform air calibration in the measured atmosphere Clock abnormalities FAIL CLOCK Abnormalities of the internal clock Make a setting of date time If such a symptom is observed repeatedly ...

Страница 44: ...s on the measured gas PH3 Set arbitrarily to 0 1 2 0 or 4 5 ppm Depends on the measured gas PH3 Set arbitrarily to 0 1 or 2 0 ppm Detection method Pump suction type Response time Within ten seconds of 0 1 ppm alarm by contact with PH3 0 3 ppm LCD display Display contents Gas name gas concentration scale bar display time battery level drawing operation and mode Power supply AA alkaline battery 2 Co...

Страница 45: ...ated in the unit of one hundredth of the volume ppm Gas concentration indicated in the unit of one millionth of the volume LEL The acronym of Lower Explosive Limit LEL refers to the lowest concentration of a combustible gas in air capable of causing explosion when ignited ...

Страница 46: ......

Отзывы: