APPENDIX B
LIFT VERSUS TUBING SIZE, FORMULATION & FLOW
ALL FLOWS SHOWN ARE ML/MIN @70°F (21.1°C)
LIFT
FEET
TUBING
SIZE
SILICONE
C-FLEX
TYGON
®
LAB
TYGON
®
LFL
NORPRENE
®
PHARMED BPT
PTFE
0
15
675
740
740
12
635
690
710
25
290
210
570
0
24
1085
1060
1040
12
730
955
1030
25
15
′
Max LIFT
560
670
0
25
750
825
785
740
12
260
625
680
690
25
15
′
Max LIFT
130
365
345
0
4 mm
30
12
29
25
20
0
6 mm
84
12
31
25
19
Notes:
1. The flows shown above were generated on an EASY-LOAD pump.
2. Average barometric pressure is 740 mm Hg.
3. Flows shown are nominal or averaged.
4. PTFE Tubing Pump is being run @ 400 rpm, which is above its recommended rpm range; this is for intermittent use
(10–15 minutes) only and will result in shorter tubing life. For continuous use, do not exceed 300 rpm. Reduce flows by 33%
for use at 300 rpm.
®