4.4.1 Setting and managing step-voltage test
Standby
status
Push ‘CLOCK’, ‘TIMER’ button to
select STEP 2.50kVSET, STEP5.00
kVSET.
Tips
Note 1: Data of the last step is held and displayed (HOLD symbol is lit up).
Note 2: When TC symbol is displayed (under the temp. compensation mode), the tester cannot conduct test under the mode
of step-voltage measurement.
Note 3: If you need to check the set voltage during measurement, just push ‘ ’ button, and the set voltage will be displayed
for about 2 seconds; At the end of measurement, push ‘ ’ button to switch between the last input voltage and the last
measured voltage.
Operation procedures:
Push ‘MEASURE’ and hold it
>1s to start step-voltage test.
Use ‘ ’,‘ ’ button to adjust
the test voltage to 5 kV.
Test voltage increases in preset steps,
and afterwards the test will be
automatically stopped.
Push ‘ ’ button, and the
set test voltage will flash.
Push ‘Enter’ button to switch to mode
of step-voltage test, and ‘Step’ symbol
will be displayed.
4.4.2 Reviewing test data at every step
Standby
status
Push ‘CLOCK’, ’TIMER’ button to
toggle between resistance and current.
Operation procedures:
Push ‘SELECT’ button, and Hold
symbol will not be displayed.
Push ‘ ’,‘ ’ button, and
select test data in each step.
Push ‘Select’ button, and
Hold symbol will flash.
Push ‘TEMP’ button to display time spent
in current step, temp., and humidity.
4.4.3 Exiting the mode of step-voltage measurement
Repeatedly push, CLOCK’ until
STEP symbol is turned off.
Operation procedures:
Push ‘ENTER’ button to return to
mode of insulation resistance test.
Standby status
Press ‘ ’