61
62
procedures
6.3.1.2 Saving data of temperature and humidity to the memory
Push ‘MEMO’ button, and ‘MEMO No.’
symbol will be displayed on screen.
Push ‘TEMP’ button,
and temperature value
will start flashing.
Standby
status
Standby
status
Push ENTER button to
save data with MEMO No.
Push ‘CLOCK’, ‘TIMER’,
‘ ’,‘ ’ to select a data no.
Clear ‘TEMP HOLD’ symbol and temperature/humidity data according to the following procedures:
Operation Procedures
Tips
Note 1: If the temperature sensor is connected, please take it off before doing anything else.
Note 2: Only the displayed temperature/humidity values are cleared in the above procedures, while data in the memory are
not cleared.
6.3.2 Clearing the indication for saving temperature/humidity data.
6.4 Communicating with a PC
1. PC can be used to make table or curve
2. Data in the memory can be uploaded to a PC, while the internal parameters can be set with a PC.
3. Relevant software shall be installed on a PC.
4. The tester cannot carry out insulation resistance test, leak current test, or voltage test when it is communicating with a PC.
5. System requirement
Operation system: Windows98, Windows2000, Windows XP, Windows Vista
Hard-disk capacity: 100 MB remaining space
Interface: USB
6. Functions of the PC software
1) Obtain the saved data from the tester.
2) Display the obtained data and log-record and make curves for the step-voltage measurement, and save test data.
3) Set tester parameters.
7. Parameters that can be set on a PC.
1) Date and time
2) Time interval for calculating PI
3) The applied time for step-voltage test
Tips
Note 1: When TC symbol is displayed, the tester returns to standby mode and humidity value is not displayed.
Note 2: With the temperature sensor being connected, humidity value cannot be displayed, even if the humidity value is held.
Note 3: When values of resistance and current are held, or when it is under the mode of step-voltage measurement,
temperature and humidity indication will be displayed and time value will be turned off after values of temperature and
humidity are entered.
Note 4: If ‘TEMP’ button is pushed with temp./humidity values flashing, system will return to standby mode.
Tips
Note: When only temperature data and humidity data are saved, they are saved as standard data. Resistance value, voltage
data and other data are saved as ---.
Push ‘CLEAR’ button, and
the temperature indication
will be changed to --- °C.
Push ‘ENTER’ button, and
the humidity indication will
start flashing.
Push ‘CLEAR’ button, and
the humidity indication will
be changed to --- %RH.
Push ‘CLEAR’ button
to return to standby mode.