Options* Description
Alarm
Priority
Factory Default
Settings
User Configurable Settings
Low
Caution
Range
Area of the display range that
provides a caution indicator.
NA
Off
Off, or 1 to 15 in steps of 1
When set to Off, alarm is disabled
Silence
Duration
Sets the amount of time that
the alarm is silenced.
NA
2 minutes
30 seconds, 1, 2, or 5 minutes
Alarm
Delay
When an alarm condition is
met, this feature delays the
audible part of an alarm.
NA
30 seconds
0, 5, 10, 15, 20, 30, or 60 seconds
* Capnography setting changes made on the Masimo device are not transferred to the Radius PCG device. Changes to
Capnography settings for the Radius PCG device are made on the device itself. See
Chapter 4: Operation
on page 27
for information about settings for the Radius PCG device.
About Parameter Information
Additional information about each parameter is available.
To access additional information about parameters:
1.
From the
Parameter Settings
screen, touch the
About
icon. The following is an example for EtCO
2
.
2.
An
About
screen appears for the selected parameter and displays information about the parameter.