Zeroing Procedure
Zeroing is recommended after 500 hours of operation or whenever an offset in gas readings is discovered. Zeroing of
the Radius PCG is performed by the following procedure:
Note:
The presence of ambient air (0% CO2) in the EMMA Airway Adapter is of crucial importance for a successful
Zeroing. Special care should be taken to avoid breathing near the Airway Adapter before or during the Zeroing
procedure.
1.
Start the Radius PCG by pressing the Power On button. See
Controls
on page 31.
2.
Make sure that a new Airway Adapter is properly fitted. See
Attaching Airway Adapter
on page 23.
3.
Press and hold down simultaneously the Power On and Alarm
Silence button until the Service Screen display the Service
code “C0” and the Service value “10”. Keep both buttons
depressed while the Service value starts "counting down" i.e.
displaying "9" - "8" - "7" etc. until "0" is displayed.
4.
When the Service value “0” is shown, zeroing of the Radius
PCG is complete.*
1
Service
Value
2
Service
Code
* The Radius PCG will return to normal measuring mode when the Service value has reached "0" or if any of the
buttons are released.