The audible alarm sound may be checked by detaching the Airway Adapter to generate a
No Adapter
alarm.
If the EtCO
2
Value is non-zero, ensure that there has not been an accumulation of CO
2
between the Radius PCG Sensor
Body and the Airway Adapter by removing and reattaching the Airway Adapter. If the EtCO
2
Value still displays a
non-zero value after this procedure, perform a Zeroing procedure. See
Zeroing Procedure
on page 60.
Power Off
The Radius PCG switches off automatically during following conditions:
•
If no breath is detected within 2 minutes from power up.
•
If no breath is detected for 2 minutes and the alarm has been silenced.
•
15 seconds after the Airway Adapter is removed.
Note:
The device will not automatically power off if there is an alarm condition other than a
No Adapter
alarm
detected (e.g.
Clogged Adapter
).
Bluetooth Setup
For Radius PCG Bluetooth Setup instructions for connection with Root, see
Pairing Radius PCG to Root
on page 67.