Model: CDU-LCD
Doc. Ref. no.:- mMP/om/601
Issue no.00.06
User’s Manual
Page 32 of 32
10.
How to get Power off time of CDU?
When CDU will power on it will log one sample with
DATA: high display range of 1.
TIME: last logged data time + periodic time saved in memory.
Ex.
Step-1: HT last log with timestamp 17:24:31
Step-2: CDU power off at 17:24:45
Step-3: CDU power on at 17:26:38
CDU power on it will punch high value of log (for HT and DP) with time 17:26:31.
11.
Why does the maximum sensor value not punch at power on?
If the Time stamp (last log time + periodic time) is greater than the power on time
of CDU, it will not punch the future time stamp.
Ex.
Step-1: Set CDU periodic Time = 2 minute.
Step-2: CDU power off at 17:30:16.
Step-3: CDU power on at 17:30:19.
Last periodic log punch with time 17:29:00, CDU power on it will not punch high value