User Manual
43
40005E (Enhanced Bargraph Indicator)
REF NO: m42B/om/101
Issue No:02
Note 3:
1)
CHANNEL Window shows sub parameter configuration name.
i.e. while setting Ethernet IP (E.Ip.), Ethenet IP 1(
E.ip.1
) will be shown on same window. And
so on. All sub parameters of Ethernet configuration will be shown as explained.
2)
E.IP.1 , E.IP.2, E.IP.3 and E.IP.4 is for setting Ethernet IP address. Value should be entered
sequencially. i.e. if Ethernet IP address of 192.168.100.190 needs to be configured , then set
value 192 to E.IP.1, set value 168 to E.IP.2, set value 100 to E.IP.3 and then set value 190 to
E.IP.4.
3)
E.msk is for setting Ethernet mask address and E. Gty is for setting Ethernet Gateway
address . Both settings must be set as explained in above note.
Table 20 Sub Parameters of Display Configuration Mode
Sub parameters of Display Configuration Mode
Channel
Window
Name
Setting name and description
Default
value
Shows
only if
Symbol
t.out
Timeout for
display back
to Run
Mode
10 to 100 Seconds
60
-
(t.oUt)
Fltr
IIR Filter to
get stable
PV reading
0 to 10 (Refer Note : 4)
0
-
(FLtr)
Note 4 :
1)
Filter is for Stability of PV data Whenever required. Filter = 0 means no filter is applied on PV
data. While Filter = 1 to 10 means Filter of Low value to High value is applied on PV data.
Table 21Sub Parameters of Logging Configuration Mode
Sub parameters of Logging Configuration Mode (Optional)
Channel
Window
Name
Setting name and description
Default value
Shows
only if
Symbol
p.log
Periodic Logging
Enable
strt
/
stop
(Strt/Stop)
0: STOP
1: START
0
-
(P.LOG)
p.l.mp
Periodic Log
channel
mapping
(selection)
yes
/
no
(YES/NO)
0 : NO
1: YES
0 (for all 2
channels)
(Refer Note 5)
-
(P.L.MP)
p.l.md
Periodic Logging
Mode
hold
/
OWLP
(HOLD/OVLP)
0: HOLD
1: OVERLAP
0
(Refer Note 5)
-
(P.L.md)
P.M.RS
Periodic Log
Memory
Reset(/Clear)
yes
/
no
(YES/NO)
0 : NO
0
(Refer Note 5)
-
(P.M.rS)