User Manual
53
40005E (Enhanced Bargraph Indicator)
REF NO: m42B/om/101
Issue No:02
By pressing MENU key, DP of last digit and last digit itself will blink to indicate that the value can be
changed. Use INC / DEC (Here Decrement key is used for shifting a digit) key to correct the displayed
reading to the desired process value and press MENU key. When the calculations are over, the new
calibration values are stored automatically and DATA window will show the calibrated value.
When user enters in calibration zero menu, CHANNEL window shows
CAL.z
(CAL.Z) message.
Pressing MENU key again CHANNEL window shows
CH-1
(CH-1) message. Press MENU key, it
shows the PV value of corresponding input for channel number 1. User can use any channel out of
maximum number of channels(2) to calibrate input type for span reading. PV display shows process
value corresponding to input sensor value with old calibration data. Feed sensor input using a
calibrator, such that process value is close to sensor‟s upper range value. Press MENU key to edit the
value. Use INC / DEC (Here Decrement key is used for shifting a digit in editing mode) key to arrive at
the desired process value. Press MENU key to register the changes. When the calculations are
over, the new calibration values are stored automatically and CHANNEL window will show the
calibrated value.
Zero and Span calibration is over
In case, the ENHANCED BARGRAPH INDICATOR cannot complete the calibration due to any
reason, it will hold previous calibration parameters. Calibration for input sensor is over.
Retransmission output calibration (Voltage/current output)(Optional):-
In Calibration mode, press INC key repeatedly, till CHANNEL window shows message
rtr.Z
(retransmission output zero calibration). By pressing MENU key, CHANNEL windows shows
rtr.1
(rtr.1) . It indicates which Retransmission output needs to be calibrated. By pressing MENU key,
CHANNEL window shows the value being outputted on Retransmission output terminals. In 40005E
there are 2 number of Analog Output(Retransmission Output) are available. Now, measure the value
using a highly accurate digital multi meter. Press MENU key to enter into editing mode. Use INC /
DEC (Here Decrement key is used for shifting a digit in editing mode) key to correct the displayed
reading to the measured value. The ENHANCED BARGRAPH INDICATOR will store zero calibration
value. Press the ESC key and then INC key to calibrate value for other Retransmission Output. Now
press ESC to come out of Retransmission Zero calibration and press INC key to calibrate
retransmission output span calibration menu.
CHANNEL window shows the message
rtr.S
(retransmission output span calibration) ) By pressing
MENU key, CHANNEL windows shows
rtr.1
(rtr.1) . It indicates which Retransmission output needs
to be calibrated. By pressing MENU key, CHANNEL window shows the value being outputted on
Retransmission output terminals. In 40005E there are 2 number of Analog Output(Retransmission
Output) are available. Now, measure the value using a highly accurate digital multi meter. Press
MENU key to enter into editing mode. Use INC / DEC (Here Decrement key is used for shifting a digit
in editing mode) key to correct the displayed reading to the measured value. The ENHANCED
BARGRAPH INDICATOR will store zero calibration value. Press the ESC key and then INC key to
calibrate value for other Retransmission Output. Now press ESC to come out of Retransmission Span
calibration. Calibration for Retransmission output is over. Press MENU key to calibrate other
parameters or press Escape key to come out to normal operation.
Note: Disable Linearization of Linear input type before doing calibration of linear input type.