•  Products with a welded plastic housing are 

not repairable. Please contact your supplier for 
replacement part.

•  This product contains hazardous voltages and 

there are no user replaceable parts inside the 
product. Never attempt to remove the cover.

•  WARNING: No modifi cation of this equipment is 

allowed. Any repair/service should be carried out 
by qualifi ed personnel who may get assistance by 
contacting the manufacturer or the manufacturer’s 
agent.

•  Products specifi ed to have automatic polarity 

protection must be switched off if a battery is 
connected with reverse polarity. The protection will 
be automatically reset when the polarity has been 
corrected.

•  In chargers specifi ed to have a replaceable fuse as 

polarity protection the fuse must be replaced if the 
battery has been connected with reverse polarity. 
When replacing the fuse; a fuse of the same type 
and rating must be used.

•  If the product is specifi ed to comply with the 

standard for Medical Electrical Equipment 
(standards based on IEC60601-1) it complies 
with some of the requirements for medical 
electrical equipment and may be used in medical 
applications and hospital environments.

•  The product must not be used in the vicinity 

of fl ammable anesthetic gases or in other 
environments with fl ammable or explosive 
atmosphere. 

•  If the product is specifi ed to comply with the 

standard for Medical Electrical Equipment 
for Home Healthcare Environment (standard 
IEC60601-1-11) it may be used in medical 
applications used in a home healthcare 
environment.

  NOTE: Products relying on safety earth for 

protection (Class I) may not be used in home 
healthcare environment unless they are 
permanently wired to the building installation:               

  Installation must only be carried out by qualifi ed 

service personnel, following the below instructions:
               

-  The protective earth conductor must be

min. 0.75 mm

2

.           

-  Connect the protective earth conductor to the 

external protective earthing system.   

-  Verify that the protective earth terminal used is 

connected to the external protective earthing 
system.             

-  Verify the integrity of the external protective 

earthing system.

•  This product converts the mains voltage to 

a safety extra low voltage. 

  The output from products applying with 2MOPP 

insulation ( model names followed by ”P”) may 
be treated as Applied Part Type B or Type BF 
according to standard EN/IEC 60601-1 and 
may come in physical contact with a patient.The 
housing of the product shall not be allowed to 
contact the patient .

•  This product must be operated in an environment 

within temperature range +5 to +40°C, humidity 
15 - 93 % RH and atmospheric pressure 70 - 
106 kPa (700 - 1060 hPa).

•  Expected service life of this product and 

accessories delivered with this product is three (3) 
years, if operated as indicated above. However; the 
guarantee times indicated in document ”TERMS 
OF SALES AND DELIVERY FOR MASCOT AS” 
apply (available at www.mascot.com).

•  Environmental parameters during transport and 

storage between uses: temperature range -25 
to +85 °C, humidity 15 - 93% RH NC and 
atmospheric pressure 70 - 106 kPa (700 - 1060 
hPa). 

•  If stored for longer periods of time the 

environmental parameters should be within 
the temperature range +5 to +35°C, humidity 
range 10 - 75% RH NC and atmospheric pressure 
70 - 106 kPa (700 - 1060 hPa) to maintain the 
products expected service life.

•  Expected shelf life of this product is one (1) year, if 

stored as indicated above.

Содержание 2040(P)

Страница 1: ...mascot no downloads usermanuals Bruksanvisning K ytt ohjeet Bedienungsanleitung Mode d emploi Manual de instrucciones Istruzioni per l uso Doc no 5170E Part No 205170 11 01 2018 MASCOT ELECTRONICS AS...

Страница 2: ...IC SHOCK READ THROUGH THESE INSTRUCTIONS PRIOR TO USING THE PRODUCT CAREFULLY FOLLOW THESE INSTRUCTIONS WHEN USING THE PRODUCT RETAIN THESE INSTRUCTIONS FOR FUTURE REFERENCE compound and is dust tight...

Страница 3: ...y should an operational error occur during use If the product has a detachable mains cord the appliance coupler may be used as a disconnect device The product is switched on by inserting the mains plu...

Страница 4: ...ay not be used in home healthcare environment unless they are permanently wired to the building installation Installation must only be carried out by qualified service personnel following the below in...

Страница 5: ...ber of cells in series If the number of cells in series in the battery pack is not known you can calculate it by dividing the voltage indicated by 3 6VDC for LI e g a 14 4VDC LI battery contains 4 Li...

Страница 6: ...10Ah Again this is just typical recommendations Please read recommendations and datasheets from battery manufacturer How to connect exchangeable AC plugs The following exchangeable AC plugs are availa...

Страница 7: ...termination level indicated on the charger The LED indication on the charger is ORANGE The battery is charged to its full capacity at the end of this step STEP 3 CHARGE COMPLETE The LED indication on...

Страница 8: ...current has dropped below a set level and the LED indication on the charger changes to YELLOW to indicate that the battery is almost fully charged and may be ready for use The constant voltage charge...

Страница 9: ...ompleted The battery is charged to its full capacity at the end of this step STEP 3 CHARGE COMPLETE The LED indication on the charger is GREEN and the battery is fully charged For Li Ion batteries the...

Страница 10: ...charge detection level or until CV timer runs out The battery is charged to its full capacity at the end of this step STEP 3 CHARGE COMPLETE The LED indication on the charger is GREEN and the battery...

Страница 11: ...oltage Cut Off Current Step 2 CV Charge max1 h Safety Timer 72 hours Step 3 Charge complete Step 0 Start Timer 10 min Step 1 CC Charge Start Current Low Current Start Up at Vbatt 3V cell Error Mode if...

Страница 12: ...8V 100mA 100mA 0 6Ah 10Ah 0 5A 21V 21V 100mA 100mA 0 7Ah 10Ah Input voltage Charge LED indicator 6 cell 21 6V 7 cell 25 2V 8 cell 28 8V 9 cell 32 4V 10 cell 36V 2240 P 2241 P 100 240Vac 50 60Hz max 0...

Страница 13: ...5Ah 10Ah Charge diagram A Technical data If not appearing in table see marking on the product Charge diagram B Input voltage Charge LED indicator 1 cell 3 6V 2 cell 7 2V 3 cell 10 8V 4 cell 14 4V 5 ce...

Страница 14: ...21V 0 6A 21V 0 6A 100mA 1 4Ah 10Ah 2440 P 2441 P 2442 P 100 240Vac 50 60Hz max 1 6A Orange CC ch Orange CV ch Yellow CV ch Green ch complete Rec batt capacity 4 5A 4 2V 4 2V 1 8A 4 2V 1 8A 300mA 4Ah...

Страница 15: ...complete Rec batt capacity 2 3A 25 2V 25 2V 1 1A 25 2V 1 1A 250mA 2 3Ah 25Ah 2 0A 29 4V 294V 1A 29 4V 1A 250mA 2Ah 25Ah 1 7A 33 6V 33 6V 0 8A 33 6V 0 8A 250mA 1 7Ah 25Ah 1 6A 37 8V 37 8V 0 8A 37 8V 0...

Страница 16: ...A 42V 42V 7A 42V 7A 3A 15Ah 300Ah Charge diagram B Technical data If not appearing in table see marking on the product Input voltage Charge LED indicator 11 cell 39 6V 12 cell 43 2V 13 cell 46 8V 14 c...

Страница 17: ...1 6A 46 2V 46 2V 0 8A 46 2V 0 8A 200mA 1 6Ah 20Ah 1 45A 50 4V 50 4V 0 6A 50 4V 0 6A 200mA 1 45Ah 20Ah 1 35A 54 6V 54 6V 0 6A 54 6V 0 6A 100mA 1 35Ah 10Ah 1 3A 58 8V 58 8V 0 6A 58 8V 0 6A 100mA 1 3Ah 1...

Страница 18: ...21 6V 7 cell 25 2V 8 cell 28 8V 9 cell 32 4V 10 cell 36V 9640 9641 220 240Vac 50 60Hz Red Orange CC ch Yellow Timer CV ch Green ch complete Rec batt capacity 1 5A 25 2V 25 2V 1 5A 4h complete 1 5Ah 6...

Страница 19: ...acity N A N A N A N A N A 2040 P 2041 P 2042 P 2140 P 100 240Vac 50 60Hz max 1 2A Orange CC ch Yellow Timer CV ch Green ch complete Rec batt capacity 1 3A 46 2V 46 2V 1 3A 4h complete 1 3Ah 52Ah 1 2A...

Страница 20: ...A 100mA 1 5Ah 60Ah 1 5A 8 4V 8 4V 1 35A 8 4V 1 35A 100mA 1 5Ah 40Ah 1 2A 12 6V 12 6V 1A 12 6V 1A 100mA 1 2Ah 48Ah 1A 16 8V 16 8V 0 8A 16 8V 0 8A 100mA 1Ah 40Ah 0 8A 21V 21V 0 65A 21V 0 65A 100mA 0 8Ah...

Отзывы: