• Products with a welded plastic housing are
not repairable. Please contact your supplier for
replacement part.
• This product contains hazardous voltages and
there are no user replaceable parts inside the
product. Never attempt to remove the cover.
• WARNING: No modifi cation of this equipment is
allowed. Any repair/service should be carried out
by qualifi ed personnel who may get assistance by
contacting the manufacturer or the manufacturer’s
agent.
• Products specifi ed to have automatic polarity
protection must be switched off if a battery is
connected with reverse polarity. The protection will
be automatically reset when the polarity has been
corrected.
• In chargers specifi ed to have a replaceable fuse as
polarity protection the fuse must be replaced if the
battery has been connected with reverse polarity.
When replacing the fuse; a fuse of the same type
and rating must be used.
• If the product is specifi ed to comply with the
standard for Medical Electrical Equipment
(standards based on IEC60601-1) it complies
with some of the requirements for medical
electrical equipment and may be used in medical
applications and hospital environments.
• The product must not be used in the vicinity
of fl ammable anesthetic gases or in other
environments with fl ammable or explosive
atmosphere.
• If the product is specifi ed to comply with the
standard for Medical Electrical Equipment
for Home Healthcare Environment (standard
IEC60601-1-11) it may be used in medical
applications used in a home healthcare
environment.
NOTE: Products relying on safety earth for
protection (Class I) may not be used in home
healthcare environment unless they are
permanently wired to the building installation:
Installation must only be carried out by qualifi ed
service personnel, following the below instructions:
- The protective earth conductor must be
min. 0.75 mm
2
.
- Connect the protective earth conductor to the
external protective earthing system.
- Verify that the protective earth terminal used is
connected to the external protective earthing
system.
- Verify the integrity of the external protective
earthing system.
• This product converts the mains voltage to
a safety extra low voltage.
The output from products applying with 2MOPP
insulation ( model names followed by ”P”) may
be treated as Applied Part Type B or Type BF
according to standard EN/IEC 60601-1 and
may come in physical contact with a patient.The
housing of the product shall not be allowed to
contact the patient .
• This product must be operated in an environment
within temperature range +5 to +40°C, humidity
15 - 93 % RH and atmospheric pressure 70 -
106 kPa (700 - 1060 hPa).
• Expected service life of this product and
accessories delivered with this product is three (3)
years, if operated as indicated above. However; the
guarantee times indicated in document ”TERMS
OF SALES AND DELIVERY FOR MASCOT AS”
apply (available at www.mascot.com).
• Environmental parameters during transport and
storage between uses: temperature range -25
to +85 °C, humidity 15 - 93% RH NC and
atmospheric pressure 70 - 106 kPa (700 - 1060
hPa).
• If stored for longer periods of time the
environmental parameters should be within
the temperature range +5 to +35°C, humidity
range 10 - 75% RH NC and atmospheric pressure
70 - 106 kPa (700 - 1060 hPa) to maintain the
products expected service life.
• Expected shelf life of this product is one (1) year, if
stored as indicated above.