MAS 640 Instruction Manual Rev. 0A
5.6.1. Autozero/Zero Tracking Options
Autozero option No.
Description:
0
Disable Autozero and Zero tracking
1
Enable Zero tracking only
2
Enable Zero tracking and Autozero
6. Setup
6.1. Entering the Setup Mode
To enter the setup mode:
●
Disconnect the indicator from the power supply.
●
Remove the instrument back cover. Move the jumper on the digital module from the position J9 to J8
[Figure 5.].
●
Turn the power on.
●
Once display shows the 'iuPER' message, move the jumper on the digital module from the position J8
to J9.
Display will light 'SET' message as shown. Now you can select any function from Main Manu Setup Functions
[6.2.] by pressing repeatedly the <PIECES> key.
Figure 5: Location of the J8, J9 jumpers on the digital module inside the instrument.
-13-