Calibration
EC
1 point slope calibration with 6 memorized solutions avail-
able: 84.0 μS/cm, 1413 μS/cm, 5.00 mS/cm, 12.88 mS/
cm, 80.0 mS/cm, 111.8 mS/cm;
1 point offset calibration: 0.00 μS/cm
NaCl 1-point, with MA 9050 calibration solution
Temp
2-point, at 0 and 50 °C (32 and 122 °F)
Temperature
Automatic or manual, from -20.0 to 120.0 °C (-4.0 to 248.0 °F)
Compensation
Temperature
Selectable from 0.00 to 6.00%/ °C (EC and TDS only)
Coefficient
TDS factor
0.40 to 0.80 (default value is 0.50)
Probe
MA 814D/1
Computer Interface
RS232/USB opto-isolated
Power supply
12 VDC power adapter
Dimensions
230 x 160 x 95 mm (9.0 x 6.3 x 3.7")
Weight
0.9 kg (2.0 lb.)
Environment
0 to 50 °C (32 to 122 °F) ; max RH 95%
Warranty
3 years
This instrument is in compliance with the CE Directives.