Instruction Manual Mi 170 Bench Meter
1 2
1 2
1 2
1 2
1 2
reading is stable.
When the reading is stable, the “READY” and “ACCEPT” tags will blink on the LCD.
Press the GLP/ACCEPT key to confirm calibration. The instrument stores the calibration
value and returns to measurement mode.
Notes
Notes
Notes
Notes
Notes:
•If the reading is too far from the expected value, the “WRONG” and “SOLUTION”
tags will blink. Calibration cannot be confirmed.
• If the temperature of the buffer is out of the 0.0 – 60.0 °C (32.0 – 140.0 °F)
temperature interval, the “WRONG” “SOLUTION” and “°C” tags will be displayed
blinking.
• If a new EC calibration is performed, the NaCl calibration is automatically cleared.
Thus, a new NaCl calibration is required.