background image

3

1. Volt - Input Terminal

This is the positive input terminal for VOLT measurements. Connection is 
made to it using the red test lead.

2. COM - Common Terminal

This is the negative (Ground) input terminal for all measurement modes 
except current. Connection is made to it using the black test lead.

3. 

Ω

Ohms input terminal

This is the positive input terminal for resistance and continuity 
measurements. Connection is made to it using the red test lead.

4. Display

The display indicates the measured value of signal, function mode, 
low battery, range, max & hold settings.

5. Function / Range selector rotary switch.

The rotary switch selects the function and desired range.

6. Max - Maximum recording mode

This function is used to measure the maximum value of AC/DC voltage, 
AC/DC current and resistance. 

7. Hold Button

Freezes the last reading displayed.

8. Clamp Jaws

Detect AC current flowing through the conductor on test.

9. Trigger

Press the lever to open the clamp jaws. When the lever is released, 
the jaws will close.

4

1.2 Unpacking and Inspection

Before unpacking the clampmeter, examine the shipping carton for any sign of
damage. Unpack and inspect the clampmeter and any associated leads for
damage. If there is any damage then consult your distributor immediately.

After removing your new Digital Clamp Meter (DCM) from its packaging, you
should have the following items:

1. Digital clampmeter
2. Test lead set (one black, one red)
3. 9-Volt battery (installed in meter)
4. Instruction manual 
5. Protective carry case 

If any of the above items are missing or are received damaged, please contact
the distributor from whom you purchased the unit.

1.3 REPLACING THE BATTERY

This meter is powered by a NEDA type 1604 or equivalent 9-volt battery. When
the meter shows the "

" sign the battery must be replaced to maintain proper

operation. Follow the procedure below to replace the battery.

1. Disconnect the test leads from any live source, turn the rotary switch to off
and remove the test leads from the input terminals.

2. The battery cover is secured to the bottom case by a screw. Using a Phillips
head screwdriver, remove the screw and then remove the battery cover.

3. Replace the battery with a new equivalent 9-volt one observing correct polarity.

4. Replace the battery cover and re-install the screw.

WARNING

To avoid electrical shock, disconnect the test leads and any input signals

before replacing the battery. Replace only with same type of battery.

5

2. SAFETY INFORMATION

The instrument complies with class II overvoltage CAT III 1000V of the 
IEC1010-1 (EN61010-1) ; UL3111-1; and CAN/CSA C22.2 #1010.1-92
standards, pollution degree 2 in accordance with IEC-664 indoor use. If the
equipment is used in a manner not specified, the protection provided by the
equipment may be impaired.

This product complies with the requirements of the following European
Community Directives: 89/336/EEC (Electromagnetic Compatibility) and
73/23/EEC (Low Voltage) as amended by 93/68/EEC (CE Marking).

Electrical noise or intense electromagnetic fields in the vicinity of the equipment
may disturb the measurement circuit. Measuring instruments will also respond to
unwanted signals that may be present within the measurement circuit. Users
should exercise care and take appropriate precautions to avoid misleading
results when taking measurements in the presence of electromagnetic
interference.

3. OPERATION

3.1 VOLTAGE MEASUREMENTS 

1. Turn off power to the circuit/device under test and discharge all capacitors.

2. Plug the black test lead into the COM input jack on the meter and connect the
test lead tip to a grounded point (the reference point for measured voltage).

3. Select the desired AC voltage range (V

), or DC voltage range (V

). If the

magnitude of the voltage to be measured is unknown, always start with the
highest range.

WARNING

To avoid possible electric shock, instrument damage and / or
equipment damage, do not attempt to take any voltage measurements 
if the voltage is above 1000V DC or 750VAC. These are the maximum
voltages that this instrument is designed to measure. The "COM"
terminal potential should not exceed 500V measured to ground.

6

4. Plug the red test lead into the VOLT input jack on the meter and connect to the
circuit where a voltage measurement is required. Voltage is always measured in
parallel across a test point.

5. Energise the circuit or device under test and make the voltage measurements.
Reduce the range setting, if set too high, until a satisfactory best resolution
reading is obtained.

6. After completing the measurement, turn off the power to the circuit / device
under test, discharge all capacitors and disconnect the meter test leads.

3.2 CURRENT MEASUREMENTS

1. Set function switch to ACA 1000A range.

2. Press the trigger to open the clamp jaws and clamp them around a conductor.
The jaws should be completely closed before taking a reading.

3. The most accurate reading will be obtained by keeping the conductor across
the centre of the clamp jaws.

WARNING

The clamp jaws are designed to take current measurements on
circuits with a maximum voltage difference of 500VAC between any
conductor and ground potential. Using the clamp jaws for current
measurements on circuits above this voltage may cause electric
shock, instrument damage and / or damage to the equipment under
test. Before measuring current, make sure that the test leads are
removed from the instrument.

The clamp jaws are overload protected up to 500VAC for up to 1
minute. Do not take current readings on circuits where the maximum
current potential is not known. Do not exceed the maximum current
that the instrument is designed to measure. 

Содержание CM54

Страница 1: ...ns and product information are available at www martindale electric co uk SYMBOLS Equipment complies with relevant EU Directives AC Alternating Current Ground Direct Current Equipment protected by Double Insulation Class II Caution refer to accompanying documents Caution risk of electric shock End of life disposal of this equipment should be in accordance with relevant EU Directives 1 CONTENTS 1 I...

Страница 2: ...ied the protection provided by the equipment may be impaired This product complies with the requirements of the following European Community Directives 89 336 EEC Electromagnetic Compatibility and 73 23 EEC Low Voltage as amended by 93 68 EEC CE Marking Electrical noise or intense electromagnetic fields in the vicinity of the equipment may disturb the measurement circuit Measuring instruments will...

Страница 3: ...he residual reading of 10 must be subtracted to obtain the actual resistance of 1500MΩ 3 4 CONTINUITY TESTING 1 Select the position by turning the rotary selector switch 2 Follow steps 2 4 as for resistance measurements 3 The clampmeter emits a tone when measuring resistance of approximately 75Ω or less After measurements have been carried out disconnect the test leads from the circuit and from th...

Страница 4: ...MΩ 0 1Ω 10Ω 10kΩ 1000kΩ 1 2 rdg 4d 1 0 rdg 3d 2 0 rdg 5d 5 rdg 10d 3 0Vdc 0 3Vdc 0 3Vdc 3 0Vdc Resolution Accuracy Open Circuit Volts G 16th Edition Testers G All in one s G Calibration Equipment G Continuity Testers G Electrician s Kits G Full Calibration Repair Service G Fuse Finders G Digital Clamp Meters G Digital Multimeters G Microwave Leakage Detectors G Motor Maintenance Equipment G Non tr...

Отзывы: