21
Range & Accuracy
Range
Min
Max
Accuracy
(% of reading ±
counts)
V Read (low)
0.000
20.000
0.015% ±2
VDC Read (high)
0.0
250.0
0.05% ±2
VAC Read (high)
0.0
250.0
0.5% ±2
V Source
0.000
20.000
0.015% ±2
mV Read
-10.00
75.00
0.015% ±2
mV Source
-10.00
75.00
0.015% ±2
mA Read
0.000
24.000
0.015% ± 2
mA Source
0.000
24.000
0.015% ± 2
CPM Source
2.0
1000.0
0.05%
Hz Source
1
1000
0.05%
KHz Source
1.0
10.0
0.25%
CPM Read
0.0
1000.0
0.05% ± 1
Hz Read
0.0
1000.0
0.05% ± 1
KHz Read
0.00
10.00
0.05% ± 1
Range
Min
Max
Accuracy
Ohms Read (low)
0.00
400.00
0.1 ohm
Ohms Read (high)
400.0
1500.0
0.5 ohm
1500.1
3200.0
1.0 ohm
Range
Min
Max
Excitation Accuracy
Current
Ohms Source (low)
5.00
400.00
0.1 to 0.5 mA
0.15 ohm
5.00
400.00
0.5 to 3 mA
0.1 ohm
Ohms Source (high)
400.0
1500.0
0.05 to 0.8 mA
0.5 ohm
1500.0
3200.0
0.05 to 0.4 mA
1.0 ohm
Thermocouple Read and Source (All errors included)
Range
Min
Max
Accuracy
J Thermocouple
-200.0
0.0
0.6°C
0.0
1200.0
0.4°C
K Thermocouple
-200.0
0.0
0.8°C
0.0
1370.0
0.5°C
T Thermocouple
-200.0
0.0
0.8°C
0.0
400.0
0.5°C
E Thermocouple
-200.0
0.0
0.5°C
0.0
950.0
0.4°C
R Thermocouple
-20
0.0
2.49°C
0
500
1.7°C
500
1750
1.3°C
S Thermocouple
-20
0
2.4°C
0
500
1.7°C
500
1750
1.4°C
B Thermocouple
600
800
2.1°C
800
1000
1.7°C
1000
1800
1.3°C