5-7
Revision A
DATA ACQUISITION ASSEMBLY AND PCB: PCB Connectors PN 800776-001B
Eagle 3000 Patient Monitor
415397-038
†
“n” in the symbol name refers to the blood pressure channel number, 1 or 2.
Table 5-4. Connector J17 (11-Pin Temperature Cable Connector)
Pin
Name
Input/
Output
Description
1
CP_+0.25VREF
O
+0.25 V reference
2
CP_TEMP1
I
Input from temperature sensor, channel 1
3
NC
4
CP_ISOAGND
O
Isolated analog ground reference
5
CP_400/700
I
Series 400/700 thermistor type
6
CP_FSHIELD
O
Cable shield
7
CP_ISOAGND
O
Isolated analog ground reference
8
CP_TEMP2
I
Input from temperature sensor, channel 2
9
NC
10
CP_TEMP_PROBE*
I
Temperature probe
11
CP_ISOGND
O
Isolated ground reference
Table 5-5. Connector J18 and J19 (11-Pin Invasive Blood Pressure Cable Connector)
Pin
Name
Input/
Output
Description
1
†
CP_B
O
Non-inverted BP transducer excitation voltage
2
†
CP_
I
Non-inverted BP transducer signal
3
NC
4
CP_ISOAGND
O
Isolated analog ground reference
5
NC
6
CP_FSHIELD
O
Cable shield
7
CP_BPn_EXCI-
O
Inverted BP transducer excitation voltage
8
CP_BPn_SIG-
I
Inverted BP transducer signal
9
NC
10
CP_BPn_ID
I/O
BP probe identification
11
NC
Содержание Eagle 3000
Страница 1: ...Eagle 3000 Patient Monitor Data Manual...
Страница 6: ...iv Revision A Table of Contents Eagle 3000 Patient Monitor 415397 038 For your notes...
Страница 18: ...1 12 Revision A INTRODUCTION Abbreviations Eagle 3000 Patient Monitor 415397 038 For your notes...
Страница 42: ...2 24 Revision A UPPER LEVEL OVERVIEW Parts List PN 800772 002A Eagle 3000 Patient Monitor 415397 038...