“F6 INP”: Display the internal counts For LED version
“F6 CAL”: Calibration for LCD Version. See below:
“F7 CAL”: Calibration
“PN”:
Press FUN Key to enter
“PN”
. Initiate password is 0000
“P1 SP”:
Set AD converting speed. Options: 1/2/3/4/5. 1 is the fastest and 5 is the slowest.
“P2 AUT”:
S
elect the “Power on “ zero maintenance, Options: 0/2/5/10/20/50/100% of the capacity.
“P3 RAN”:
Select the manually zero maintenance, Options: 2/ 4/10/20/50/100% of the capacity.
“P4 AZ”:
Select the auto zero maintenance, Options: 0.5d/1d/2d/4d.
“d” means division.
“P5 DEC”:
Set the decimal point. Options: 0/0.0/0.00/0.000/0.0000.
“P6 CAP”:
Set the capacity.
“P7 INC”:
Set division increment. Options: 1/2/5/10/20/50.
Make sure enter all P5,P6 ,P7 and set parameters in order
“P8 CAL”:
Calibration
After finishing all necessary settings above, the indicator is ready for calibration. (Note: If the scale has settings
already done you can directly calibrate it.) Press FUN Key when displaying
“P8 CAL”. Display will show
“noLoAd”. Empty the platform by removing the weight if any. When platform is empty and STABLE indicator is on
then press FUN Key to confirm. Display shows
“XXXXXX”. Set it to be the weight of the weigher you want to put
on. Press FUN Key to confirm, and it
will show “LoAd”. Put the calibration weigher on the platform and press FUN
Key after the STABLE indicator is on. Indicator will go to weighing state. The calibration is finished.
“P9 INT”:
Initialize the memory to ex-work state.
“P10 RS”:
Communication mode.
This option is used for the RS232C serial port m
ode setup. Options: “RS Cot” is continuous Data Transmission
for using Remote Display o
r External Display, “RS Prt” is used for the Serial Printer. “RS Aut” is used for
“transmission after every weighing”.
“P11 FL”:
Weight adjustment for acceleration of gravity.
“P12 KB”:
When
“on”, keyboard is pressed with sound.
“P13 UT”:
Unit setting. Lb, U1 or U2 is optional, which could be
“ON” or “OFF” through pressing the FUN Key
and TARE Key. When U1/U2 is
“ON”, the conversion coefficient can be set through pressing UNIT Key while
self-checking. When U1/U2 is
“OFF”, it could not be set unless entering the P13 UT menu and set it “ON”.
“P14 Pn”:
Press
“FUN” to display “P1 ----” set the passwords and press “FUN”. It display “P2 ----” and set
passwords again to confirm. When successful ,it will show
“Done”
SECTION 5 RS-232 OUTPUT
Specifications: RS-232 output of weighing data, 1200/2400/4800/9600/19200 bps,1 stop bit, 8 data bits, No
Parity.
Mode 1: Continue output