MARA OPERATOR’S MANUAL
MARA OPERATOR’S MANUAL
APPENDIX A:
TECHNICAL DESCRIPTION
Input Rating
110-240V ~50/60HZ 9A
Operating Parameters
Transport and Storage
Duty Cycle
Continuous with Intermittent operation (3 min on / 7 min off) during water
vapor delivery.
Cleaning the
Console
and Cartridge
WARNING:
Always turn off and unplug the
Console
before cleaning.
Follow the procedures approved by your institution or use a validated
infection control procedure.
• Thoroughly wipe all surfaces of the Console, Cartridge and power
cord with a mild cleaning solution (listed below) and a damp cloth.
Do not clean the
Console
or Cartridge with abrasive compounds,
solvents, or other materials that could scratch the panels or damage
the
Console
. Use common hospital cleaning solutions, such as:
• 70% Alcohol
• 3% Hydrogen Peroxide Solution
• 10% Bleach Solution
Do not allow cleaning fluids to enter the
Console
or Cartridge.
Cartridge must be fully dry before use with Water Vapor Probe.
Do not attempt to sterilize the
Console
or Cartridge.
Ambient Temperature Range
Relative Humidity
Atmospheric Pressure
Warm-up Time
10º to 35º C (50º to 95ºF)
30% to 75%, noncondensing
700 to 1060 hPa
If the Console was transported or stored at temperatures
outside the Transport and Storage ambient temperature
range, allow at least one hour for the Console to reach
room temperature before use.
Ambient Temperature Range
Relative Humidity
Atmospheric Pressure
Duration of Storage
2º to 60º C (36º to 140ºF)
25% to 85%, condensing
500 to 1060 hPa
If the Console was stored for longer than one year, the
Console should be tested and calibrated before use.