SECTION 5
TROUBLESHOOTING
The following is a list of common
malfunctions, alarms and alert codes.
Note:
As indicated with # in the table
below, please follow these instructions
when removal of the Water Vapor
Probe is indicated: To remove the
Water Vapor Probe, press the red
Treatment Interrupt Button, twist to
release, choose either “Clear” or “Quit,”
and then disconnect the Water
Vapor Probe.
If the issue persists with
multiple
Water
Vapor Probes, turn off
the Console and contact an AEGEA
Medical representative.
Note:
If an error(s) continuously
occurs while a Water Vapor Probe is
connected to the Console, turn off
the Console’s power and disconnect
the Water Vapor Probe. Power on the
Console and continue using a new
Water Vapor Probe.
For service contact information, refer
to back of Operator’s Manual.
Malfunctions
Should the Console malfunction,
return to AEGEA Medical Inc. for
service. Do not attempt to repair
or service if troubleshooting
is ineffective.
SECTION 4
AFTER THE PROCEDURE
Powering Off and Storing the
Console:
1. Disconnect the Water Vapor Probe
handle from the Cartridge and
disconnect the Syringe from the
Syringe Port. Dispose per your
institution’s guidelines
(image 10)
.
2. Wrap the Cartridge Cable around
the storage features on the side of
the Console and insert the Cartridge
into the Cradle
(image 10)
.
3. Unlock and lower the pole for the
saline bag
(image 10)
.
4. Press the Front Panel Power Button
on front of Console to turn the
unit off.
5. Disconnect the power cord from
the back of the Console. Store the
power cord with the Console.
Image 10