Millenium HX RO System
3030226 Rev. D
6-4
Troubleshooting
6.3
Product Water Quality Alarm
1.
Product Quality Shutdown during normal run.
a.
The Product water Quality set point is adjustable from 1 to 50 TDS (1-100
µS). Verify that your set point is correctly set.
6.4
This Unit Must Stand Upright Alarm
1.
This Unit Must Stand Upright shutdown.
a.
Verify the Unit is in the upright position.
b.
With the unit disconnected from electricity, verify that the tilt sensor is
closed using an ohmmeter when in the upright position. If faulty, replace tilt
sensor.
Содержание MILLENIUM HX
Страница 1: ...3030226 Rev D Mar Cor Purification MILLENIUM HX REVERSE OSMOSIS UNIT Operation and Maintenance Manual...
Страница 2: ......
Страница 4: ...This page intentionally left blank...
Страница 10: ...Millenium HX RO System 3030226 Rev D 1 2 General Information This page intentionally left blank...
Страница 18: ...Millenium HX RO System 3030226 Rev D 1 10 General Information NOTES...
Страница 33: ...Millenium HX RO System 3030226 Rev D 2 15 Installation And Initial System Start Up NOTES...
Страница 35: ...Millenium HX RO System 3030226 Rev D 3 1 System Operation Millenium HX RO System CHAPTER 3 SYSTEM OPERATION...
Страница 36: ...Millenium HX RO System 3030226 Rev D 3 2 System Operation This page intentionally left blank...
Страница 52: ...Millenium HX RO System 3030226 Rev D 3 18 System Operation NOTES...
Страница 54: ...Millenium HX RO System 3030226 Rev D 4 2 Screen Operation And Specifications This page intentionally left blank...
Страница 88: ...Millenium HX RO System 3030226 Rev D 5 2 Routine Maintenance This page intentionally left blank...
Страница 94: ...Millenium HX RO System 3030226 Rev D 5 8 Routine Maintenance Tubing Connections...
Страница 117: ...Millenium HX RO System 3030226 Rev D 6 1 Troubleshooting Millenium HX RO System CHAPTER 6 TROUBLESHOOTING...
Страница 118: ...Millenium HX RO System 3030226 Rev D 6 2 Troubleshooting This page intentionally left blank...
Страница 128: ...Millenium HX RO System 3030226 Rev D 6 12 Troubleshooting NOTES...
Страница 129: ...Millenium HX RO System 3030226 Rev D 7 1 Drawings Millenium HX RO System CHAPTER 7 DRAWINGS...
Страница 130: ...Millenium HX RO System 3030226 Rev D 7 2 Drawings This page intentionally left blank...
Страница 141: ...Millenium HX RO System 3030226 Rev D 7 13 Drawings 7 2 Flow Schematics 7 2 1 Run Mode...
Страница 142: ...Millenium HX RO System 3030226 Rev D 7 14 Drawings 7 2 2 Divert to Drain...
Страница 143: ...Millenium HX RO System 3030226 Rev D 7 15 Drawings 7 2 3 Heat Chemical Purge...
Страница 144: ...Millenium HX RO System 3030226 Rev D 7 16 Drawings 7 2 4 Heat Chemical Recirculation...
Страница 145: ...Millenium HX RO System 3030226 Rev D 7 17 Drawings 7 3 Electrical Diagram...
Страница 147: ...Millenium HX RO System 3030226 Rev D 7 19 Drawings NOTES...
Страница 148: ...Millenium HX RO System 3030226 Rev D 7 20 Drawings This page intentionally left blank...
Страница 149: ...Millenium HX RO System 3030226 Rev D 8 1 Spare Parts Millenium HX RO System CHAPTER 8 SPARE PARTS...
Страница 150: ...Millenium HX RO System 3030226 Rev D 8 2 Spare Parts This page intentionally left blank...
Страница 155: ...Millenium HX RO System 3030226 Rev D 8 7 Spare Parts NOTES...
Страница 156: ...Millenium HX RO System 3030226 Rev D 8 8 Spare Parts This page intentionally left blank...
Страница 157: ......