2
Operating instructions VENTA Multi Care 16_ 5752 2434
GA 5752 2434 EN 13
20
Personal safety notes
Basic safety notes
W
A R N I N G
!
Injury hazard!
Patient may be endangered as a result of incorrect use.
Follow the operating instructions for all accessories.
W
A R N I N G
!
Injury hazard!
Electrical devices (e. g. cellphones, radios, magnetic resonance tomographs) can
interfere with the functioning of the product when used near the product.
Electrical devices which can interfere with the functioning of the product may not
be used near the product.
Please observe the information concerning electromagnetic compatibility (EMC)
(emissions and resistance to interference) contained in the Technical specifica-
tions.
Adhere to those specifications when using electrical devices and respond
properly in the event of effects on the device or the product.
W
A R N I N G
!
Infection hazard resulting from improper handling!
Applicable rules for hygiene shall be observed in order to avoid infection or bac-
terial contamination when extracting and disposing of secretions. Observe the
purpose for which the bacteria filter is intended. Use sterile catheters only during
extraction and ensure that the patient is not injured during the procedure. Always
wear gloves during the procedure.
W
A R N I N G
!
Infection hazard resulting from using the wrong hydrophobic bacteria filter!
The septic fluid suction pump may be used only with a hydrophobic bacteria filter
suitable for the specific application. This is to protect it against contaminants
which might be present in the form of particles or aerosols in the gas being drawn
in and to prevent contamination of the air nearby.
Always use a hydrophobic bacteria filter appropriate to the particular extraction
situation.
W
A R N I N G
!
Infection hazard due to using no hydrophobic bacteria filter or the wrong type!
Secretions will enter the septic fluid suction pump during extraction.
Do not continue using the septic fluid suction pump. Clean and disinfect the septic
fluid suction pump and have it repaired by a shop authorised by MAQUET to do
so.
Содержание 5752 2345
Страница 1: ...OPERATING INSTRUCTIONS MEDAP SEPTIC FLUID SUCTION PUMP VENTA MULTI CARE 16 GA 5752 2434 EN 13 ...
Страница 6: ...Operating instructions VENTA Multi Care 16_ 5752 2434 GA 5752 2434 EN 13 iv ...
Страница 68: ...11 Operating instructions VENTA Multi Care 16_ 5752 2434 GA 5752 2434 EN 13 62 Notes Spare parts ...
Страница 71: ......