Mai ntenance
63
Step
Action
1
Move the pipette partway into the Clean Location.
2
Grasp the rotating assembly of the pipette with one hand, and remove the steel retain-
ing nut where the paddle stem meets the pipette with your other hand.
3
Let the retaining nut drop into the Clean Location.
Move the pipette to the Load Location
To move the pipette to the Load Location, follow the instruction video and/or the soft-
ware guided workflow integrated in the MicroCal PEAQ-ITC Control Software.
The software guided workflow is also shown in the following table.
Step
Action
1
Make sure that the Load Location does not contain a microcentrifuge tube.
2
Move the pipette to the Load Location.
Note
: It should not be necessary, but have your hand ready to receive the syringe glass.
3
Click
Next
.
Result
: The instrument will move the plunger down and press out the syringe glass.
Remove the syringe glass
To remove the syringe glass, follow the instruction video and/or the software guided work-
flow integrated in the MicroCal PEAQ-ITC Control Software.
The software guided workflow is also shown in the following table.