Installati on
35
Site requirements
Introduction
This section outlines the site requirements for MicroCal PEAQ-ITC.
Bench space
The MicroCal PEAQ-ITC system (instrument and Controller PC) requires about 1.2 meters
of bench space (depth approximately 64 cm). For more details, See Site requirements.
!
WARNING! Access to power switch and power cord with plug
. Do not
block access to the power switch and power cord. The power switch must
always be easy to access. The power cord with plug must always be easy to dis-
connect.
Ambient environment
Environmental fluctuations may cause subtle performance problems.
The following should be avoided:
l
Strong draft
l
Room temperature fluctuations
l
Intense sunlight
l
Vibrations
l
Strong electrical or magnetic fields (as may be produced by an NMR, microwave
oven, large motors or refrigeration units)
The following table shows the environmental operating requirements.
Specification
Requirement
Temperature
10°C to 28°C
Humidity
0% to 70% RH, non-condensing
Atmospheric pressure 700 to 1060 hPa