Installati on
36
Electrical power
The power source (100-240 VAC) should be free from voltage fluctuations, harmonic dis-
tortions, power dips and spikes.
!
WARNING! Electrical power
. Provide proper electrical power to the instru-
ment. This should be 100-240 VAC, 50/60 Hz, observing any local electrical
safety requirements. Only power cords supplied with the instrument should be
used.
!
WARNING! Protective ground
. The product must always be connected to a
grounded power outlet.
The AC power line should be dedicated to MicroCal PEAQ-ITC and should not share
power with additional equipment.
Although the power filtering in the MicroCal PEAQ-ITC instrument is adequate for most
laboratory environments, some disturbances may affect the performance of the instru-
ment and it may be necessary to have the AC power source evaluated (see the following
table) or install a power conditioner.
Because power source problems can be manifested in many different ways, it is not pos-
sible to recommend a power conditioner for all situations. It is recommended that you
test a power conditioner, at your location, before you purchase it.
If you believe you are experiencing power source related problems, please contact a Mal-
vern Instruments field engineer.
The following table show the power supply requirements.
Specification Requirement
Voltage
100-240 VAC (power adapter), 24 VDC (power supply to the instrument)
Frequency
50/60 Hz
Power
150 W