Bedpan Washer Disinfector - Energy Saver
Detergent Model With Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 21
Page 21
14 September 2017
Appendix A
–
Faults
Diagram B1 (below) shows an example of fault mode display (Fault 01 indicated).
Diagram B1
Following is a more detailed list of fault conditions and most likely causes.
Fault 01
Door Open Fault
–
Not Confirmed Door Closed
(i.e. Door switch still indicates open)
Cause
Door limit switch stuck out/broken
Fault 02
Wash /
Disinfection
Cycle Time Out
(taking too long to complete cycle)
Cause
Steam leak
Final rinse solenoid stuck open
Faulty temperature sensor hot water tank
Blown element leg
Steam condensate leaking cold water into chamber
Faulty temperature sensor
–
chamber
Fault 03
Faulty Hot Water Probes
(high indicates water, but low does not indicate water)
Cause
Faulty water level probe
Fault 04
Door Motor Activation Time Out
(taking too long to close or open door)
Cause
Faulty door actuator
Door axle bar sticking
Fault 06
Hot Water Tank Fill Time Exceeded
(high level not reached during pre-fill, resulting in time out)
Cause
Faulty fill solenoid
Faulty water level probe
Cold water tank empty or not filling quick enough
Low water pressure