![Malmet Energy Saver ES910 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/malmet/energy-saver-es910/energy-saver-es910_operation-maintenance-and-installation-manual_3921055016.webp)
Bedpan Washer Disinfector - Energy Saver
Detergent Model With Hands Free Operation
Operation, Maintenance and Installation Manual
Issue 21
Page 16
14 September 2017
Note: Regardless of element combinations 15 seconds before the end of the cycle ALL elements
will turn off. This is to overcome the effects of overshoot in the steam generator producing
steam at the end of the disinfection cycle.
18.
The maximum time allowable for the disinfection cycle is 10 minutes. If this time is exceeded then the
machine will go to fault and elements turn off.
19.
After the disinfection cycle is complete the cooling solenoid will turn on for 15 seconds. This will cool
down the surface temperature of the utensils and make them safe to handle.
The display shows cool
. This water is direct from the mains.
20.
At the end of the cool down rinse,
the display shows end
. This will be displayed until the door is re-
opened.
21.
If not used the unit steam generator temperature will be kept at the temperature of 85°C-90°C for a
period of 15 minutes. If the unit is still not used it will enter into idle mode and the elements will only
come on at 50°C and rise to the temperature of 90°C repeatedly until the cycle is started again.