8
Care and advice on operation
Makita SK103P
en
GB
F
I
E
P
NL
DK
S
N
FIN
J
CN
ROK
PL
HR
H
RUS
CZ
GR
LT
LV
EST
SLO
SK
M
TR
Checking the accuracy of the vertical
plumbing of the SK103P:
Place the Lock switch
3
in the "Unlocked" (
)
position.
Checking the upper plumbing point:
Checking the lower plumbing point:
Set up the laser on its tripod or wall mount bracket near point
A1 at a minimum distance of 1.5 m from point B1. The hori-
zontal laser is aligned in direction 1. Mark the laser dots A1
and B1 with a pin.
Rotate the instrument by 180° so that it points in the opposite
direction 2 to direction 1. Adjust the instrument so that the
laser beam hits point A1 exactly. If point B2 is no further than
2 mm away from point B1, then the SK103P is within toler-
ance.
Should your SK103P be outside of the specified
tolerance, please contact an authorised dealership of
Makita.
Display notices
Falling below or exceeding the permissible temperature
range:
The laser switches off and all symbols flash.
Outside of the self-levelling range:
The laser beam is flashing.
Pendulum locked:
The laser beam is not levelled and the Lock symbol
4
lights
up.
Care and advice on operation
Do not immerse the instrument in water. Wipe off dirt with a
damp soft cloth. Do not use aggressive cleaning agents or
solvents. Treat the instrument with the same care that you
would apply to binoculars or a camera.
Dropping or violent shaking of the instrument may damage
it. Check the instrument for any damage before using it.
Check the levelling accuracy of the instrument regularly.
Transport
To safely transport the instrument, set the Lock switch
3
to
"Locked" (
).
1
B1
A1
B2
1.5 m
max. 2 mm
2
1
B1
A1
B2
1.5 m
max. 2 mm
2