90
The recirculating fan
is not working during
the operation.
Warning pops up. System
shut off automatically. Press
enter to turn off the
warning.
Contact the service
department of Major
Science for maintenance.
The platform is not
working due to the
belt malfunctioning.
Warning pops up. System
shut off automatically. Press
enter to turn off the
warning.
Contact the service
department of Major
Science for repair.
The transparent
shield is not closed
while operating.
Warning pops up in
“program” mode. System
shut off automatically. Press
enter to turn off the
warning/Restart
automatically after door is
closed.
Close the shield to start up
the operation again.
The device is under
dangerous condition
due to over shaking.
Icon turns red. System
slows down
the shaking
speed automatically to the
range which is indicated safe
by g-sensor.
Adjust the shaking speed to
lower point/ Adjust the foot
stands to make sure the
device is placed stably/
Contact
your
local
distributor to adjust the
sensitivity of g-sensor if
needed.
The power was cut
off during the
operation.
The full-screen alert will
show up along with
5-second beeps when user
restart the device, and the
system will automatically
start the operation from the
time point that the device
was shut down.
Press
during the
beeps to cancel the
operation and start a new
one. Or press
after
the beeps to remove the
sign and continue the
original operation.
Содержание Winpact SI-200
Страница 2: ...2...
Страница 14: ...14 PT 100 Sensor Port External Temp Port Spare Installation Hole Fan...
Страница 16: ...16 Loosen the flathead screws on both ends of the rack Lift up the rack from the bottom plate Flathead Screw...
Страница 27: ...27 A maximum of 9 microplate holders can be installed if the same type of holders is used...
Страница 37: ...37 Features of Control Panel LCD Screen Key Pad 1 2 3 5 4 6 7 8...
Страница 61: ...61 Step8 When the installation complete press Close to exit the installation dialogue...
Страница 66: ...66 B Status Column B 1 B 2 B 3 B 4...