Operation Manual PILOT TEST
8504660
Rev.
3
33
06/11/2019
5.5.10 Reset to Default
Figure 30
Figure 31
Returns device settings to factory defaults by
pressing the
Enter
key (Figure 30). Press
any other key to return. Once entered a
secondary menu is displayed to confirm
selection (Figure 31).
5.6 Troubleshooting
If your device is no longer working properly, see Table 13.
Table 13 Troubleshooting
PROBLEM
REASON
SUGGESTION
Lamps do not
light up
Lack of
power
supply
•
Is the power switch on?
•
Is the power cord plugged in correctly in the power
connector?
•
Is the wall outlet working?
No sound
from
headphone
Disturbed
connection
•
Is the headphone cable plugged in correctly to the
socket?
•
Is the lead loose or defective?
•
Is the cord cracked or bent?
No sound
from monitor
phone
Disturbed
connection
•
Is the cable plugged in correctly to the socket?
•
Is the lead loose or defective?
NOTE
: Monitor is only available for PILOT TEST, not
Audiometry.
Printing is not
possible
Disturbed
connection,
lack of paper
•
Is the printer cable connected to the device and printer?
•
Is the printer on?
•
Is there paper in the printer?
NOTE
: If there are any problems that you cannot solve yourself, please contact your local
distributor.