No modification of this equipment is allowed.
Equipment is not user repairable. Repairs must be
performed by a qualified service representative only. No
modifications of the equipment are allowed by anyone other
than a qualified MAICO representative.
Modification of the equipment could be hazardous.
No part of the equipment can be serviced or maintained
while in use with the patient.
Do not drop or otherwise cause undue impact to this device.
If the device is dropped or otherwise damaged, return it to
the manufacturer for repair and/or calibration. Do not use
the device if any damage is suspected.
Calibration of the device: The audiometer and the
headphone complement each other and share the same
serial number (i.e. MA7663252). Therefore, the device shall
not be used with any other headphone prior to recalibration.
Recalibration also needs to be conducted, when a defected
headphone is replaced.
Uncalibrated devices may lead to faulty measurements and
sometimes even damage the hearing of the examinee.
2.7 Electrical Safety and Measuring Security
This icon indicates that patient applied parts of the device
conform to IEC 60601-1 Type B requirements.
The device is not intended for operation in areas with an
explosion hazard. Do NOT use the device in a highly
oxygen-enriched environment, such as a hyperbaric
chamber, oxygen tent, etc. If the device is not used switch
it off and disconnect it from the power supply.
Never short-circuit the terminals.
In order to maintain a high level of safety and to ensure the
device works properly, it is necessary to have the device and
its power supply checked according to the medical electrical
safety standard IEC 60601-1 by a qualified service technician
at least once a year. For more information see Section 3.2.
The use of non-calibrated devices can lead to incorrect test
results and is not advisable.