Table 15 Reference Values for Stimulus Calibration
REFERENCE VALUES FOR STIMULUS CALIBRATION
Fre-
quency
[Hz]
Reference equivalent threshold sound pressure level [RETSPL, dB re. 20
Pa]
according to:
PTB Report 2009,
DTU Report 2010
Coupler IEC 60318-3
PTB Report 2018,
DTU Report 2018
Coupler IEC 60318-1
DD45
DD65 v2
125
47.5
30.5
250
27.0
17.0
500
13.0
8.0
750
6.5
5.5
1000
6.0
4.5
1500
8.0
2.5
2000
8.0
2.5
3000
8.0
2.0
4000
9.0
9.5
6000
20.5
21.0
8000
12.0
21.0
Table 16 Frequencies and Maximum Intensities: AC (Air Condition) dB HL
TRANSDUCER MAXIMUM HEARING LEVELS
Frequency
[Hz]
Intensities [dB HL]
DD45
DD65 v2
Tone
Tone
125
70
70
250
90
90
500
100
100
750
100
100
1000
100
100
1500
100
100
2000
100
100
3000
100
100
4000
100
100
6000
100
85
8000
90
70