Operation Manual MA 25/MA 25e
MA 27/MA 27e
8104818 Rev. 5
23
29/06/2021
5.7.3.3 Screening
A hearing screening utilizes a
Pass
or
Refer
result and is used to determine if further
testing is required as a hearing problem may exist. Patients are typically screened at
a level of
20 dB HL
at
500 Hz
,
1000 Hz
,
2000 Hz
, and
4000Hz
in
each ear
. If a patient
hears all the tones in each ear, the result would be considered a
Pass
. Failure to hear
even one of the tones in either ear would result in a
Refer
.
NOTE
: This is an example of one screening protocol. Each state may have their own
screening protocol. Contact your state health department for guidelines in your area.
5.7.4 Automatic Threshold (Hughson-Westlake, MA 25e/MA 27e Only)
In addition to traditional manual testing, the MA 25e/MA 27e incorporates a Hughson-
Westlake patient controlled automatic threshold test complying with ISO8253. When the test
is completed the results are easily recalled from the internal memory of the MA 25e/MA 27e.
Hughson-Westlake is a procedure used to determine pure tone thresholds. The
MA 25e/MA27e utilizes this procedure to perform an automatic pure tone test procedure.
Threshold is defined as 2 out of 3 (or 3 out of 5) correct responses obtained at a certain
level in a 10 dB decrease and 5 dB increase procedure. The device will re-test 1000 Hz
before moving to the next ear or ending the test.
Prior to testing, the following instructions should be given. "
You will now hear a variety
of tones with various loudness levels, please push the response switch when you
hear a tone and release the button when you no longer hear it
“.
The patient response
can only be recorded during tone presentation.
The test frequencies will start at 1000 Hz and continue through those frequencies
activated within the settings.
To start the automatic test, press the
Frequency Hz
dial. This will change the Function
Key list to select HW with F4. For key selection see Table 8 in Section 5.5.2.
5.8 Tone Setup Menu
To access the
Tone Setup menu
press
F1
and
F4
simultaneously for 2-3
seconds. Once in the Menu (Figure 12), the different Setup options are listed and can
be entered using the function keys or the
Frequency Hz
10 for further explanation.
Table 9 Explanation of Function Keys in Setup Menu
Function key
Label
Description
F1
Change
To change highlighted setting.
F2
↑
To browse up in the setup menu.
F3
↓
To browse down in the setup menu.
F4
Save
To save setting and go back to previous screen.