Automatic Hughson-Westlake Test (HW)
The MA 25e and MA 27e incorporate the
automatic Hughson-Westlake Test (HW)
.
The automation of this test is configured in the Hughson-Westlake test setup menu.
Press
Change
to access the
Hughson-Westlake Tests setup
menu. Press
Change
again to enter the single setting options. Press Save to return to the main setup menu.
Table 11 Hughson-Westlake Test
Setup Menu
Description
HW Familiarization
To select if the patient shall be trained with a familiarization
test (
On
), or not (
Off
).
HW Condition
The HW test can be automated to confirm
2
– 3
(2 out of 3)
or
3
– 5
(3 out of 5) correct answers before moving to the
next frequency.
HW Frequencies…
The HW allows for test frequencies to be deactivated
separate from the manual audiometric test process. Press
Change
to toggle between the 7 frequencies that can be set
to
On
or
Off
:
125 Hz, 250 Hz; 750 Hz; 1500 Hz, 3000 Hz,
6000 Hz, 8000 Hz
. Press
Save
to return to the main
Hughson-Westlake Tests Setup
Menu.
HW On Time
Press
Change
to set the stimulus on time to 1 or 2 seconds.
HW Random time
off
Press
Change
to set the random time. The random time can
be set between 0 and 1.6 seconds.
HW Lower limit
Press
Change
to set the lower screening limit and define
when to move on to the next frequency. The lower limit can
be set between -10 and 20 dB.