5.10.2.6 Minimum Value (MIN VALUE)
Figure 60
With the MIN VALUE setting (Figure 60) the user is allowed to
include the minimum OAE response amplitude to the
PASS/REFER criterion for a single frequency. When the MIN
VALUE setting is set between -10 dB SPL and + 5 dB SPL the
OAE response for the frequency under test needs to be at an
equal or higher level than the MIN VALUE set to get a PASS
result. If MIN VALUE is set to OFF, the PASS decision is based
on the SNR alone for the current frequency measured.
Default MIN VALUE setting is -10 dB SPL.
CHANGE
to select a value between -
10 dB SPL
to
5 dB
SPL
or to set it
OFF
.
NEXT
to proceed to the next screen.
5.10.2.7 Clock Mode
Figure 61
The
CLOCK MODE
menu allows the user to change the clock
from a 24 hour mode to a 12 hour mode (Figure 61).
CHANGE
to select 1
2 Hour
or
24 Hour
.
NEXT
to proceed to the next screen.
5.10.2.8 Graph Style
Figure 62
The user can select between two options of displaying the result.
(Figure 62). The SNR graph displays the Signal-to-noise ratio as
a function of frequency. The value graph view shows absolute
OAE levels and noise levels as a function of frequency.
Section 5.7.1 Calibration and Test
CHANGE
to select
GRAPH SNR
or
GRAPH VALUE
.
NEXT
to proceed to the next screen.