3.4 Cleaning the probe tip
In order to secure correct impedance measurements it is important to make sure that the
probe system is kept clean at all times. Therefore please follow the below illustr
instructions on how to remove e.g. cerumen from the small acoustic and air pressure channels
of probe tip.
easyTymp®
Operating Manual
12
3.4 Cleaning the probe tip
In order to secure correct impedance measurements it is important to make sure that the
probe system is kept clean at all times. Therefore please follow the below illustr
instructions on how to remove e.g. cerumen from the small acoustic and air pressure channels
Never clean the probe tip while the tip is still fitted on
the probe.
Unscrew the probe cap of the probe by turning counter
clock direction.
Take the probe tip out of the probe.
Take the cleaning tool apart to
find the thin brush and
thin rigid plastic cord.
Use the plastic cord or brush to push contamination out
of the probe tip. Always enter the probe tip from the
rear to avoid accumulation of contamination inside the
vents.
Place the probe tip back onto
the probe. Make sure that
the different holes fit in the corresponding cavities.
Screw the probe cap back on the probe. Finger force
will tighten the screw sufficiently. Never use tools to fix
the probe cap!
If any blockage or damage occurs to the sealing gasket,
the probe system can only be serviced by MAICO.
Operating Manual
In order to secure correct impedance measurements it is important to make sure that the
probe system is kept clean at all times. Therefore please follow the below illustr
ated
instructions on how to remove e.g. cerumen from the small acoustic and air pressure channels
Never clean the probe tip while the tip is still fitted on
Unscrew the probe cap of the probe by turning counter
find the thin brush and
Use the plastic cord or brush to push contamination out
of the probe tip. Always enter the probe tip from the
rear to avoid accumulation of contamination inside the
the probe. Make sure that
the different holes fit in the corresponding cavities.
Screw the probe cap back on the probe. Finger force
will tighten the screw sufficiently. Never use tools to fix
If any blockage or damage occurs to the sealing gasket,
the probe system can only be serviced by MAICO.