9
Mahr GmbH • Multimar 844 T
5. Setting with measuring standards (e.g. setting ring, combination of gage
blocks, master, etc.)
Comparative measurement
-
Position the Multimar 844 T on the measuring standard
-
With light pendular motion determine the position of the reversal point (Min. and/or Max
value) and here set the indicating instrument to "0".
Absolute measurement
-
Enter the PRESET value of the indicating instrument corresponding to the measuring
standard
-
Position the Multimar 844 T on the measuring standard
-
With light pendular motion determine the position of the reversal point (Min. and/or Max
value) and here set the indicating instrument to the default PRESET value
Please note:
To conduct absolute measurement a digital indicator with a dynamic measuring function
and a PRESET is required. The following indicating instruments are recommended:
Model Range
Product Types
Digital Indicators
MarCator 1087 BR, 1087 BRi
Содержание Multimar 844 T
Страница 16: ...16 Mahr GmbH Multimar 844 T ...