Mahr GmbH • Millimar C1202
6.2
System 2
–
Press the ON/OFF key
=>
Menu appears.
– Press
Ó
,
▼
or
▼
and
to select
System 2
.
6.2.1 Updating module N 1700
!
The firmware file for the N 1700
module must first be copied to a
Micro SD card using a PC. The card
is then inserted into the C 1202.
–
Select the
Flash module
submenu
using the
▼
or
▲
key.
–
Select using
4
key or exit submenu
using the
Ó
key.
– Use
▼
or
▲
to select the firmware
file for the N 1700 module in use, or
exit the submenu using the
Ó
key.
!
Please check that the right firmware
file is saved on the Micro SD card.
If the wrong firmware is sent to
the N 1700 module, it will stop
working and will have to be sent to
Mahr Service.
–
Confirm selection by pressing key
4