Mahr GmbH • Millimar C1202
The instrument is most accurate after approx. 30 minutes of operation. To perform a
reliable customer calibration using a probe, the probe’s electrical zero point should be
positioned at the calibration zero point using the probe’s raw value display. (see point
6.1.1.).
This is because the inductive probe measures most accurately at the electrical zero
point. The probe’s electrical zero point should be set on a 1.000 mm gage block for
a customer calibration, like the one in this example, where a 2.000 mm gage block is
used for the MAX value and the measuring plate represents the MIN value.
6.1.2.1.2 Performing customer calibration
–
Select the
Calibration
submenu
using the
▼
or
▲
key.
–
Select using
4
–
Select the
Calibration C1
submenu
using the
▼
or
▲
key.
–
Select using
4
–
Select Perform customer calibration
using the
▼
or
▲
key.
–
Select using
4
key or exit submenu
using the
Ó
key.
– Use
▼
to select the value for
Calibration min
.
–
Use the
4
key to select the
Calibration min
submenu.