Maxuum III Digital Electronic Indicator 2239028 Rev D
8.2.1 Protocol
NOTE: When using serial output cable 2239036 or 2239038
Protocol for com setting is 8-N-1 9600 Baud.
The character strings will be in one of the following formats:
1start bit
2 stop bits
7 bits ASCII character
2400 baud
No parity bits
r
r
1.99mm
d
s
<SP>
DD ? DC
DB DA M
t
<SP>
<LF>
r
r
0.100in
d s <SP> ? DD DC DB DA I t <SP> <LF>
r
r
1.00mm
d
s
<SP>
DD ? DC
DB DA M
t
<SP>
<LF>
r
r
0.040in
d s <SP> ?
0 DC DB DA
I t <SP> <LF>
r
r
0.199mm
d s <SP> ? DD DC DB <SP> M t <SP> <LF>
r
r
0.0199in
d s <SP> ?
0 DC DB DA
I t <SP> <LF>
DA, DB, DC, DD
Numeric Values
<SP>
= ASCII space character
<LF>
= ASCII Line Feed character
I
= indication of inch measurements
M
= indication of mm measurements
s
= indication of measurement polarity
d
= indication of dynamic display mode. Can assume one of the following:
When
HOLD
function is
ACTIVE
A
= indication of actual mode
M
= indication of maximum mode
L
= indication of minimum mode
T
= indication of TIR mode
When
HOLD
function is
INACTIVE
a
= indication of actual mode
m
= indication of maximum mode
l
= indication of minimum mode
t
= indication of TIR mode
<SP>
= if the standard product mode is active
t
= indication of possible over or under limit condition
(active in standard mode only). Can assume one of the following:
0
= exceeds upper tolerance limit
U
= exceeds lower tolerance limit
<SP>
= if the enhanced product mode is active.